4.1 Testing the EPC 10 USB with the Model Circuit
23
In a similar way as you explored the
C-slow
compensation, you could now have a closer look into the Rs com-
pensation. Turn the compensation on by setting an appropriate compensation speed, 2, 10 or 100
µs
(9), and
gradually increase the percentage of compensation from 0 to 95% by clicking and dragging the mouse upwards
(10). As soon as you are overcompensating the series resistance typical oscillations will occur in the
Oscilloscope
.
Series Resistance Compensation is a more complicated topic and is therefore treated in more detail in chapter 7.1
on page 43.
The steps listed above can be automatically executed by clicking the
Whole-Cell
button or pressing the key ”3” on
the numerical keypad (
Amplifier Window
has to be active). This will execute the following protocol (see
Protocol
Editor
) that sets the right
Gain
and does a
C-slow
compensation with reasonable values:
3 : WHOLE-CELL
E
Mode
3
; Whole Cell
E
Gain:
11
; 10 mV/pA
E
CSlow:
50.00pF
;
E
RSeries:
20.0MOhm
;
E
AutoCSlow:
;
E
AutoCSlow:
;
E
Beep
;
4.1.5
”Whole-Cell” - Current Clamp Recording
If
C-slow
has been compensated so far, switch into the current clamp mode by selecting
C-Clamp
from the
Recording Mode
pop up. This automatically selects the voltage monitor (Vmon) as the active channel displayed
in the
Oscilloscope
. Note, that the unit of the
Test Pulse
amplitude changes from ”mV” to ”pA” as soon as you
switch from voltage clamp (VC) into current clamp (CC) mode.
Patchmaster
uses two different amplitudes for
VC and CC modes, therefore the
Test Pulse
is set to ”0 pA” initially. Now you need to inject current into the
circuitry, 100 pA should be a reasonable value. The current injection will charge the ”membrane” of the ”model
cell” at a time constant
τ
=
R
m
∗
X
C
m
= 500
M
Ω
∗
22
pF
= 11
ms
to a final value of
V
max
=
R
m
∗
X
I
= 500
M
Ω
∗
100
pA
= 50
mV
Due to the slower time constant compared with voltage clamp conditions it takes much longer to reach
Vmax
,
therefore you should increase the duration of the
Test Pulse
to a more appropriate value of 100 ms.
Note:
In contrast to voltage clamp conditions, were
τ
is proportional to the access- or series resistance
(
R
s
) of the pipette, in current clamp experiments
τ
depends on the membrane resistance (
R
m
).
The normal setting of the
Oscilloscope
scales the voltage monitor at 250 mV per division. You should therefore
increase the gain of the
Oscilloscope
to 16 which scales the display to be 16 mV per division. Please remember
that the
Oscilloscope
gain is different from the amplifier
Gain
and only scales the display, not the acquisition of
data. Using a very high
Oscilloscope
gain together with a low amplifier
Gain
allows you to determine the digital
resolution of the analog-to-digital converter.
Note:
The
500
M
Ω
setting of the model circuit is not a good method for testing the fast clamp speed
of the
EPC 10 USB
due to the long time constant of 11 ms which the amplifier can easily follow. If
you want to have a better estimation of the amplifier’s speed under current clamp conditions you should
do the same test as above with the
10
M
Ω
setting. This results in a much shorter ”membrane” time
constant of only
60
µs
.
http://www.heka.com
Summary of Contents for EPC 10 USB
Page 1: ...Hardware Manual Version 2 8 EPC 10 USB Computer controlled Patch Clamp Amplifier...
Page 6: ......
Page 10: ...4 Introduction http www heka com...
Page 16: ...10 Description of the Hardware http www heka com...
Page 22: ...16 Installation http www heka com...
Page 32: ...26 Verifying and Testing the EPC 10 USB http www heka com...
Page 44: ...38 The control software http www heka com...
Page 48: ...42 Operating Modes http www heka com...
Page 54: ...48 Compensation Procedures http www heka com...
Page 58: ...52 Patch Clamp Setup http www heka com...
Page 64: ...58 Using the Patch Clamp http www heka com...
Page 74: ...68 Appendix II Probe Adapters http www heka com...
Page 76: ...70 Appendix III S Probe http www heka com...
Page 81: ......