6
2.4
Instrumentarm and instrument table
1.
With the instrument arm wrapped up in plastic film, carefully pull wires
and tubes free.
2.
Place a metal washer (1) on the top of the console as indicated. Next, place
the ball bearing (2) and then a metal washer (3). Lastly, place the Teflon
washer (4) as shown in the diagram below.
Parts are delivered in a plastic bag.
3.
Lift the instrument arm and place it gently in the consol top for arm.
The instrument arm needs to fit tightly to the surface.
4.
Thread the water and air hoses through the hole (5).
The green hose (water) is mounted on the water distribution device (6).
The blue hose (air) is mounted on the air distribution device (7).
5.
Mount the Return air device (7a) and connect the yellow hose, Dim. Ø4
between (7b) and (7c).
6.
Attach the yellow return air hose, Dim. Ø6 to the return air device for
connection (8).
7.
Mount the ground wire with screws in the consol (9).
8.
The wire for the instrument tabel is mounted with a socket onto the main
board (10)placed in the top right corner. Be aware that the socket is placed
proper.
9.
If there is another 2 pole cable from the instrument table, please connect
this cable to the main board screw terminal (11).
10.
Mount and tighten the friction break screw (incl. the washer and break
cylinder) in the consol (12) to the dentist’s preference.
11.
Carefully un wrap the instrument table and the instrument arm.
Summary of Contents for UnicLine 5D
Page 7: ...7...
Page 13: ...13 3 Final Test To be completed by authorized technician...
Page 15: ...15 To be completed by authorized technician and the Dentist...
Page 16: ...16 5 Installation plan and floor design 500 min 440 min 440 min 540 465 735 500...
Page 18: ...18 5 1 Drilling diagram Please use the enclosed diagram when placing the unit in the clinic...
Page 19: ...Baldersh j 38 DK 2635 Ish j Denmark Tel 45 4332 0990 Fax 45 4332 0980 www heka dental dk...