background image

 

 

 

10 

2.10 

Connections 

1. 

Rinse the water supply and ventilate the air supply completely, before 
connecting the unit (for at least for 5 min.). 

2. 

Connect the green tube (water) to the tap. Use the supplied fittings 
and the water filter input slot. Open the water supply. Make sure you 
tighten and inspect all connections for any leaks. 

3. 

Connect the blue tube (air) to the tap. Use the supplied fittings and 
the air filter input slot. Open the air supply. Make sure you tighten 
and inspect all connections for any leaks. 

4. 

Connect the suction system from the unit to the suction system in the 
Clinic. The Unit can be delivered with either wet- or dry suction. 

5. 

Connect the waste tube from the spittoon and cup to the supplied  

 

fittings. Connect the air filter ventilation tube. 

6. 

Connect the waste fitting to the clinic waste pipe. 

7. 

Mount the suction motor cable to the power distribution board 
terminal labelled ‘Aspir’. 

8. 

Connect main power to the power distribution board terminal labelled 
‘Mains’. 

 

Summary of Contents for UnicLine 5D

Page 1: ...3 Mounting instructions UnicLine 5D UnicLine patient chair Dental light KEEP THIS MANUAL WITH UNIT AT ALL TIMES Manuals for OEM equipment are included in the shipping boxes Installation Service and M...

Page 2: ...3 Console 4 2 4 Instrumentarm and instrument table 6 2 5 Foot control 8 2 6 Operating lamp 8 2 7 X ray machine 8 2 8 Call system 8 2 9 Patient chair 9 2 10 Connections 10 2 13 Adjustment 11 3 Final Te...

Page 3: ...ng parts UnicLine patient chair Dental light Unpack the unit and chair with caution Remove the bubble plastic around the unit console and demount the sides of the console by turning the 2 large screws...

Page 4: ...r must be fitted to the water inlet pipe Hard water can very quickly ruin a dental unit 4 Air pressure 4 5 8 bar Balltap with R 3 8 inner thread Top 5 cm above the floor 5 Drain outside diameter 35 mm...

Page 5: ...a level on the arm s contact face on the console where the arm set is to be mounted and adjust the balance screws with the enclosed tool until the console is levelled The console is to be as close to...

Page 6: ...istribution device 6 The blue hose air is mounted on the air distribution device 7 5 Mount the Return air device 7a and connect the yellow hose Dim 4 between 7b and 7c 6 Attach the yellow return air h...

Page 7: ...7...

Page 8: ...power distribution board terminal labelled Lamps If the unit is provided with a turn on turn off function please connect the lamp cable to the circuit labelled ON OFF 2 7 X ray machine 1 Mount the ad...

Page 9: ...unt the chair on the floor as shown in the installation plan 4 Mount the armrest If ordered 5 The Cable from the chair leads through the relief connecting piece page 7 Connect it to the power distribu...

Page 10: ...and the air filter input slot Open the air supply Make sure you tighten and inspect all connections for any leaks 4 Connect the suction system from the unit to the suction system in the Clinic The Un...

Page 11: ...yringe with water until all air has left the system 4 Adjust the water pressure to 2 Bar and push button 1 down again 5 Pull bottom 2 on the air filter up and adjust air pressure to 4 5 Bar 6 Mount th...

Page 12: ...the dentist s desired quantity using the knobs under the instrument table 11 Position any hand held instruments the dentist normally uses together with the surgery equipment 12 Dismount cover on the b...

Page 13: ...13 3 Final Test To be completed by authorized technician...

Page 14: ...plate for the console Attention The installation card must be completed and returned to Heka Dental A S by an authorized service engineer The two annual service cards must be completed by an authoriz...

Page 15: ...15 To be completed by authorized technician and the Dentist...

Page 16: ...16 5 Installation plan and floor design 500 min 440 min 440 min 540 465 735 500...

Page 17: ...leitung f r Saugmaschine 5x1 5 Conducteur pour pompe aspirante 2x1 5 Styreledning til sug 2x1 5 F Water outflow without thread 32 mm Wasserabfluss ohne Gewinde DN40 Evacuation d eau sans filetage 32 m...

Page 18: ...18 5 1 Drilling diagram Please use the enclosed diagram when placing the unit in the clinic...

Page 19: ...Baldersh j 38 DK 2635 Ish j Denmark Tel 45 4332 0990 Fax 45 4332 0980 www heka dental dk...

Reviews: