12
8.
Mount the two suction hoses ensuring the larger hose is placed away
from the chair.
9.
If the unit is provided with an automatic suction pipe cleaning system,
fill the container with cleaning fluid.
10.
Mount all instruments on the instrument bridge. Test each instrument
function. Adjust the spray quantity to the dentist’s desired quantity
using the knobs under the instrument table.
11.
Position any hand held instruments the dentist normally uses together
with the surgery equipment.
12.
Dismount cover on the backside (5). Dismount the instrument joint (6)
(mounted with a magnet). Lastly, dismount cover (7) on the arm.
13.
Loosen the break screws (8) for the springs. Adjust the springs.
Tighten the break screws for the springs (9) and (10).
14.
Remount all covers.
15.
Power the patient chair on.
16.
Program any position the dentist prefers.