2.13
Adjustment
1.
Turn the unit on.
2.
Pull knob (1) on water filter up and adjust the water pressure to 0.5
Bar.
3.
Activate the syringe with water until all air has left the system.
4.
Adjust the water pressure to 2 Bar and push button (1) down again.
5.
Pull bottom (2) on the air filter up and adjust air pressure to 4.5 Bar.
6.
Mount the spittoon rinse pipe (3) and the cup fill pipe (4).
7.
Place a cup in the cup holder. Program the cup fill – and the spittoon
rinse time.