background image

 

 

 

nose approximately 3/4 inch (20 mm) above the exhaust port and sniff the exhaust. 

14. If the exhaust does not have a readily detectable level  of odor, repeat step 13.   If the exhaust 

has a readily detectable odor, press the 

“READ” key to record

 

the corrected RDL reading 

level along with GNSS location and timestamp.  The display may briefly show the 
RDL icon and the test reading.  The 

Readily Detectable Level 

should be detectable by 

the average 

person’s sense of smell. 

15. If optionally enabled via OdorSeer Configuration, the instrument can now be used to 

perform a 

Relative Odor Intensity

 (ROI) test.  The LCD will 

display “ROI x.xx” where “ROI” 

will flash and “x.xx” is the live reading.  Adjust the Flow Metering Valve to obtain the pre-
determined target reading on the LCD.  If configured via OdorSeer, the pre-determined 
target reading may be viewed by pressing and holding the “VIEW” key.  Upon release, the 
display will return to “ROI x.xx”. 

16. Wait about  30 seconds, position the nose approximately 3/4 inch (20 mm) above the 

exhaust port and sniff the exhaust.  Rate the relative odor intensity on a scale from 1 

– 5 

where the relative odor intensity is: 

 

 

– absent, 

 

 

– barely detectable, 

 

 

– readily detectable, 

 

 

– strong or 

 

 

– very strong or obnoxious. 

17. Enter the rated odor relative intensity in the instrument by using the 

“↑” or “↓” keys to 

increment or decrement the count until the desired value is shown on the display.   The 
display count will change by one for each 

“↑” or “↓” key press and alternate between “ROI” 

and the count value.  For example, after 5 

“↑” key presses “5 0.80” would indicate a rating 

of 5 relative to a test reading of 0.80 % gas in air. 

18. P

ress the “READ” key to record the ROI rating along with ROI target value, actual 

corrected % gas concentration, GNSS location and timestamp. 

 

**NOTE**  

The rating must be entered before the “READ” key will be active. 

 
19. The display will show the RDL icon and test reading for 5 seconds.  Should the RDL 

reading exceed the configurable (via OdorSeer) RDL High Alarm, the display will show 
“RDL” “HIGH” for 5 seconds along with an audible beep from the instrument to alert the 
operator of the RDL test result.  The ODORATOR 2 will now go into PURGE mode and the 
display will show the PURGE icon, a flashing “ * “ and the live reading.  The Flow Metering 
Valve must be closed while in PURGE mode 

and the live reading should be “0.00”.  In 

PURGE mode, the last set of readings may be reviewed by pressing 

either the “↑” or “↓” 

keys to scroll through the last ZERO, TDL, RDL and optionally ROI values.  After the 
PURGE mode times out the ODORATOR 2 will automatically power off and enter SLEEP 
mode 

(“ZZZZZ” will briefly display).  During PURGE mode the LCD shows a flashing 

asterisk 

“ * ” to allow the operator to abort PURGE mode and re-enter  CONDITIONING 

mode for additional tests 

by pressing the “ * “ key.  The operator may also choose to press 

and hold the “PWR” key to turn the ODORATOR 2 “OFF” after sufficiently purging the flow 
system by entering SLEEP mode 

(“ZZZZZ” will briefly display). 

Summary of Contents for Odorator 2

Page 1: ...ODORATOR 2TM Operator s Manual All Models...

Page 2: ...onsent of Heath Consultants Incorporated Heath Consultants Incorporated operates under a continual product improvement program and reserves the right to make improvements and or changes without prior...

Page 3: ...Heath Consultants Incorporated 9030 Monroe Road Houston TX 77061 Tel 713 844 1300 www heathus com ODORATOR 2 Operator s Manual Models 1 00 Methane 0 50 Methane 0 15 Methane 0 44 Propane...

Page 4: ...ocedure 7 Bluetooth 10 ODORATOR 2 Operator Controls and Connectors 10 Figure 1 ODORATOR 2 Top Panel Layout 10 Figure 2 ODORATOR 2 Rear Case View 11 Battery Charging 11 Battery Pack Replacement 12 CHAP...

Page 5: ...to show the actual breadcrumb trail used The electronics and the speed of the blower are calibrated so that the concentration of the gas air sample at the exhaust port agrees with the LCD reading at 2...

Page 6: ...S MORE THAN FOUR PSIG COULD RUPTURE THE LOW PRESSURE REGULATOR DIAPHRAM VENT THE GAS OUT OF THE INSTRUMENTAND CAUSE INSTRUMENTFAILURE THE RECOMMENDED INLET SUPPLY PRESSURE IS 7 WATER COLUMN WHEN OPERA...

Page 7: ...THE RISK OF IGNITION OF A FLAMMABLE OR EXPLOSIVE ATMOSPHERE ANY CABLE POWER or USB MUST ONLY BE CONNECTED IN A LOCATION KNOWN TO BE NON HAZARDOUS DO NOT OBSTRUCT OR BLOCK THE INTAKE OR EXHAUST PORTS...

Page 8: ...nt for Type 3R testing UL50 CSA 22 2 No 94 1 12th edition Battery Pack Run Time Internal rechargeable Li Ion 3 weeks typical run time Battery Charger Power Supply External in line 110 240 VAC 1 6A 50...

Page 9: ...of affecting their sense of smell Proper use of the ODORATOR 2 requires an operator with a normal sense of smell List of Contents The following items are included with the ODORATOR 2 1 ODORATOR 2 ins...

Page 10: ...ibration 12 CAL OK 13 A live reading display in Gas unit The actual battery capacity will be shown on the bar graph on the left side of the display The turn on sequence will generally take 20 30 secon...

Page 11: ...ed and the ODORATOR 2 turned OFF Odor Threshold Level and Relative Odor Intensity Tests Procedure 1 Prior to attaching the sample hose sniff the sample hose and verify no odor is detected If odor is d...

Page 12: ...ten only 7 Wait about 30 seconds and then position the nose approximately 3 4 inch 20 mm above the exhaust port and sniff the exhaust The exhaust must be un odorized If odorant is detected the ODORATO...

Page 13: ...shown on the display The display count will change by one for each or key press and alternate between ROI and the count value For example after 5 key presses 5 0 80 would indicate a rating of 5 relati...

Page 14: ...ndle is included for sample hose and fittings storage Bluetooth NOTE Bluetooth hardware is installed and configured but connectivity is not supported A future firmware and OdorSeer OdorSeer Remote rel...

Page 15: ...leep The ODORATOR 2 must be recharged in this case prior to use NOTE In order to obtain full battery capacity charge the battery pack when the ambient temperature is above 50 F 10 C and below 85 oF 30...

Page 16: ...the top of the cover slightly out then up to remove the cover 5 Squeeze the cable connector s lock to disengage and unplug the battery from its extension cable 6 Plug in the new battery pack The inst...

Page 17: ...forms on pages 16 and 17 are provided to document the Periodic Tests as a courtesy only Your Company s procedures will define how records are kept and in what format Calibration For regulatory complia...

Page 18: ...always detected Contaminated sample hose or flow system Change tubing RMA The RMA process may be found on page 18 of this manual Parts List The following list specifies ODORATOR 2 parts and accessori...

Page 19: ...or this test may be found on page 13 of this manual The forms on pages 16 and 17 are provided for your convenience They should remain in the manual and be kept in a safe location as a permanent record...

Page 20: ...16 Periodic Test Records Test Date Results Action Operator Name...

Page 21: ...17 Periodic Test Records Test Date Results Action Operator Name...

Page 22: ...er than a Heath warranty repair location Batteries and damage from battery leakage and all expendable items such as filters and tubing are excluded from this warranty Heath s responsibility is express...

Page 23: ...air form requests information such as complete shipping and billing addresses instrument or product name model number and serial numbers Also included will be a brief description of the problem you ar...

Page 24: ...061 Model Odorator 2 Has been demonstrated to be in compliance with the European EMC directive 2004 108 EC FCC 47 CFR Part 15 and Industry Canada ICES 003 and is in accordance with the following stand...

Page 25: ...Park Way Monongahela PA 15063 Phone 724 242 3145 Fax 724 872 3206 Southwest Region Heath Consultants Incorporated 9030 Monroe Road Houston TX 77061 Phone 713 844 1300 Fax 713 844 1309 Western Region H...

Page 26: ...22...

Page 27: ...__________________________________________________________________ ___________________________________________________________________________ _________________________________________________________...

Page 28: ...Heath Consultants Incorporated 9030 Monroe Road Houston TX 77061 Tel 713 844 1300 www heathus com...

Reviews: