background image

 

 

 

 
4.  Open the gas supply outlet valve and the ODORATOR 2 Flow Metering Valve fully for 

about 10 seconds to purge the air out of the sample hose and condition the instrument.  Do 
not sniff the exhaust while conditioning. 

5.  It is recommended to perform a Flow Self-Test at the start of each day of testing or when a 

test reading is suspect.  With the Flow Metering Valve fully open, to perform a Flow Self-
Test

, press the “SELF TEST” key.  If no gas flow is detected, “OPEN” “VALVE” and the 

ERROR icon will be displayed for 5 seconds before returning to live readings. Check the 
gas supply and all valves prior to repeating the Flow Self-

Test by again pressing the “SELF 

TEST” key.  To pass Flow Self-Test the flow must be sufficient to produce a reading greater 
than or equal to the calibration target value (typically 1.00 %) and then SELF TEST OK and 
the reading will be displayed for 5 seconds before returning to live readings.  If the Flow 
Self-Test reading is less than the calibration target value the Flow Self-Test will fail - 
ERROR and the reading will be displayed for 5 seconds before returning to live readings.  
Check the gas supply and all valves prior to repeating the Flow Self-Test by again pressing 
the “SELF TEST” key.  It is recommended to not use the ODORATOR 2 for odor level 
testing if the Flow Self-Test fails.  All Flow Self-Test results are recorded.  Fully close the 
Flow Metering Valve after performing a Flow Self-Test, wait about 30 seconds and do not 
sniff the exhaust port during a Flow Self-Test. 

6.  Close the Flow Metering Valve.  

Do not force, finger tighten only!

 

7.  Wait about 30 seconds and then position the nose approximately 3/4 inch (20 mm) above 

the exhaust port and sniff the exhaust. The exhaust must be un-odorized.  If odorant is 
detected, the ODORATOR 2 must not be used for odor level testing. 

8.  Press the 

“ZERO” key to automatically zero the reading.  If gas flow is detected, “CLOSE” 

“VALVE” will be displayed for 5 seconds before returning to live readings.  Verify all valves 
are closed before repeating zeroing by again pressing the “ZERO” key. While zeroing, the 
ZERO icon will flash.  When complete, the new zero reading along with the non-flashing 
ZERO and VIEW icons will be held in the display.

 

9. 

Press the “READ” key and the LCD will show “TDL” for performing a 

Threshold Detection 

Level

 (TDL) test.

 

10. To perform a TDL test, slowly open the Flow Metering Valve counter-clockwise by 1/4 

– ½ 

turn, wait about 30 seconds, position the nose approximately 3/4 inch (20 mm) above the 
exhaust port and sniff the exhaust.  Your upper lip may  lightly touch the front edge of the 
exhaust port.

 

11. If no odor is detected in the exhaust, repeat step 10.

 

 

**NOTE**

   

The operator  should  frequently  pause  when  sniffing  for  odorant by moving the 

nose away from the instrument,  breathing fresh air and then continue testing.  This 
precaution  is  necessary  because  the  sense  of  smell  fatigues rapidly during this type of 
test. 

 
12. The 

first faint smell 

of odor is called the TDL level.    Considerable variation will exist 

among individuals in detecting this threshold level.  When detected, 

press the “READ” key 

to record the corrected TDL reading level along with GNSS location and timestamp. 

13. 

The LCD will show “RDL” for performing a 

Readily Detectable Level

 (RDL) test.  Slowly 

open the Flow Metering Valve another 1/4 

– 1/2 turn, wait about 30 seconds, position the 

Summary of Contents for Odorator 2

Page 1: ...ODORATOR 2TM Operator s Manual All Models...

Page 2: ...onsent of Heath Consultants Incorporated Heath Consultants Incorporated operates under a continual product improvement program and reserves the right to make improvements and or changes without prior...

Page 3: ...Heath Consultants Incorporated 9030 Monroe Road Houston TX 77061 Tel 713 844 1300 www heathus com ODORATOR 2 Operator s Manual Models 1 00 Methane 0 50 Methane 0 15 Methane 0 44 Propane...

Page 4: ...ocedure 7 Bluetooth 10 ODORATOR 2 Operator Controls and Connectors 10 Figure 1 ODORATOR 2 Top Panel Layout 10 Figure 2 ODORATOR 2 Rear Case View 11 Battery Charging 11 Battery Pack Replacement 12 CHAP...

Page 5: ...to show the actual breadcrumb trail used The electronics and the speed of the blower are calibrated so that the concentration of the gas air sample at the exhaust port agrees with the LCD reading at 2...

Page 6: ...S MORE THAN FOUR PSIG COULD RUPTURE THE LOW PRESSURE REGULATOR DIAPHRAM VENT THE GAS OUT OF THE INSTRUMENTAND CAUSE INSTRUMENTFAILURE THE RECOMMENDED INLET SUPPLY PRESSURE IS 7 WATER COLUMN WHEN OPERA...

Page 7: ...THE RISK OF IGNITION OF A FLAMMABLE OR EXPLOSIVE ATMOSPHERE ANY CABLE POWER or USB MUST ONLY BE CONNECTED IN A LOCATION KNOWN TO BE NON HAZARDOUS DO NOT OBSTRUCT OR BLOCK THE INTAKE OR EXHAUST PORTS...

Page 8: ...nt for Type 3R testing UL50 CSA 22 2 No 94 1 12th edition Battery Pack Run Time Internal rechargeable Li Ion 3 weeks typical run time Battery Charger Power Supply External in line 110 240 VAC 1 6A 50...

Page 9: ...of affecting their sense of smell Proper use of the ODORATOR 2 requires an operator with a normal sense of smell List of Contents The following items are included with the ODORATOR 2 1 ODORATOR 2 ins...

Page 10: ...ibration 12 CAL OK 13 A live reading display in Gas unit The actual battery capacity will be shown on the bar graph on the left side of the display The turn on sequence will generally take 20 30 secon...

Page 11: ...ed and the ODORATOR 2 turned OFF Odor Threshold Level and Relative Odor Intensity Tests Procedure 1 Prior to attaching the sample hose sniff the sample hose and verify no odor is detected If odor is d...

Page 12: ...ten only 7 Wait about 30 seconds and then position the nose approximately 3 4 inch 20 mm above the exhaust port and sniff the exhaust The exhaust must be un odorized If odorant is detected the ODORATO...

Page 13: ...shown on the display The display count will change by one for each or key press and alternate between ROI and the count value For example after 5 key presses 5 0 80 would indicate a rating of 5 relati...

Page 14: ...ndle is included for sample hose and fittings storage Bluetooth NOTE Bluetooth hardware is installed and configured but connectivity is not supported A future firmware and OdorSeer OdorSeer Remote rel...

Page 15: ...leep The ODORATOR 2 must be recharged in this case prior to use NOTE In order to obtain full battery capacity charge the battery pack when the ambient temperature is above 50 F 10 C and below 85 oF 30...

Page 16: ...the top of the cover slightly out then up to remove the cover 5 Squeeze the cable connector s lock to disengage and unplug the battery from its extension cable 6 Plug in the new battery pack The inst...

Page 17: ...forms on pages 16 and 17 are provided to document the Periodic Tests as a courtesy only Your Company s procedures will define how records are kept and in what format Calibration For regulatory complia...

Page 18: ...always detected Contaminated sample hose or flow system Change tubing RMA The RMA process may be found on page 18 of this manual Parts List The following list specifies ODORATOR 2 parts and accessori...

Page 19: ...or this test may be found on page 13 of this manual The forms on pages 16 and 17 are provided for your convenience They should remain in the manual and be kept in a safe location as a permanent record...

Page 20: ...16 Periodic Test Records Test Date Results Action Operator Name...

Page 21: ...17 Periodic Test Records Test Date Results Action Operator Name...

Page 22: ...er than a Heath warranty repair location Batteries and damage from battery leakage and all expendable items such as filters and tubing are excluded from this warranty Heath s responsibility is express...

Page 23: ...air form requests information such as complete shipping and billing addresses instrument or product name model number and serial numbers Also included will be a brief description of the problem you ar...

Page 24: ...061 Model Odorator 2 Has been demonstrated to be in compliance with the European EMC directive 2004 108 EC FCC 47 CFR Part 15 and Industry Canada ICES 003 and is in accordance with the following stand...

Page 25: ...Park Way Monongahela PA 15063 Phone 724 242 3145 Fax 724 872 3206 Southwest Region Heath Consultants Incorporated 9030 Monroe Road Houston TX 77061 Phone 713 844 1300 Fax 713 844 1309 Western Region H...

Page 26: ...22...

Page 27: ...__________________________________________________________________ ___________________________________________________________________________ _________________________________________________________...

Page 28: ...Heath Consultants Incorporated 9030 Monroe Road Houston TX 77061 Tel 713 844 1300 www heathus com...

Reviews: