CAUTION AND WARNING 

 

To prevent injury and damage to your scale, please follow these instructions very carefully: 

 

 

Do not transport the scale while tray is loaded. 

 

For accurate weighing the scale must be placed on a flat, stable surface. 

 

For accurate weighing, verify proper operation according to the procedure described in this manual before each use. 

 

Do not use in the presence of flammable materials. 

 

Operating at other voltages and frequencies than specified could damage the equipment. 

 

If the “LOW BAT” indicator activates, for accurate weighing replace the batteries or plug the scale into a power source as soon 
as possible. 

 

Only manufacturer’s qualified personnel should service the scale. 

 

This equipment has been tested and found to comply with the EMC limits for the Medical Device Directive 93/42/EEC (EN 

55011 Class A and EN 60601-1-2).  These limits are designed to provide reasonable protection against harmful interference in 

a typical medical installation. The equipment generates, uses and may radiate radio frequency energy and, if not installed and 

used in accordance with these instructions, may cause harmful interference to other devices in the vicinity. However, there is 

no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference with 

other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to eliminate the 
interference by one or more of the following measures: 

 

Reorient or relocate the receiving device. 

 

Increase the separation between the equipment. 

 

Connect the scale’s adapter into an outlet on a circuit different from that to which the other device(s) is connected. 

 

Consult the manufacturer at (+1) 800-638-3722 or (+1) 708-377-0600, or a field service technician for help. 

 

DEFINITION OF SYMBOLS  

 

 

 

 

 

 

 

 

 

 

 

 

553KL SCALE SPECIFICATIONS  

 

Capacity and Resolution: 

44 lb x 0.5 oz / 20 kg x 0.01 kg 

Power Requirements: 

6-AA cell 1.5V batteries (INCLUDED) or optional Adapter Model ADPT30  

120-240V AC – 9V DC 50-60Hz.  OPTIONAL ADAPTER NOT INCLUDED. 

Environmental: 

Operating Temperatures: 50°F to 104°F / 10°C to 40° C 

Atmospheric Pressure: 500hPa to 1060hPa 

Humidity: 10% to 90% RH 

Tray Dimensions: 

25” x 12-¾”  /  63.5 cm x 32.4 cm 

Scale Dimensions: 

25” x 16-¼ x 3-¾”  /  63.5 cm x 41.3 cm x 9.5 cm 

 

STORAGE SPECIFICATIONS   

Storage Temperatures: 

32°F to 122°F / 0°C to 50°C 

Humidity: 

10% - 90% RH 

Atmospheric Pressure: 

500hPa – 1060hPa 

 

 

 

553KL SCALE CONNECTIVITY INFORMATION: 

 

This scale is designed to connect to other devices via a USB port, to allow acquisition  

of weight from the scale, to help close a source of potential medical errors caused by  

incorrectly copying and then recording a weight taken.  This scale is EMR-EHR compatible.   

For more information regarding how to connect this scale to your computer, a monitor or  

other device please contact Health o meter® Professional Technical Support at 800-638-3722. 

 

 

 

 

 

 

*WARNING:

    

For use with Power Adaptor Model UE 15WCP-090050SPA/ADPT30 only.  
Complies with C22.2 No. 601.1-M90 and UL Std No. 60601-1. Type BF 

applied parts. Internally powered

 

Equipment is not suitable for use in the 

presence of flammable anesthetic mixture with air nor with oxygen or 
nitrous oxide.

 

AC to DC power supply 
(9V DC, 200mA) 

 

DC Connector Polarity 

Temperature range 

(50°F to 104°F / 10°C to 50°C) 

Humidity range 

(10% - 90% RH) 

Atmospheric pressure range 

(500hPa – 1060hPa) 

Attention, Consult Accompanying Documents 

Type BF Applied Parts 

AC to DC power supply 

USB 

Summary of Contents for 553KL

Page 1: ...RIC SCALE BALANZA ELECTRONICA PEDI TRICA BALANCE ELECTRONIQUE PEDIATRIQUE 44 lb x 0 5 oz 20 kg x 0 01 kg Rev 20100611 USER INSTRUCTIONS INSTRUCCI NES DE USO MODE D EMPLOI 232791 IEC 60601 1 C22 2 No 6...

Page 2: ...manuel et le garder port e de main pour pouvoir le consulter tout moment TABLE OF CONTENTS ENGLISH 2 ESPA OL 8 FRAN AIS 14 CAUTION AND WARNING 3 SYMBOLS AND SPECIFICATIONS FOR THIS SCALE 3 CONNECTIVI...

Page 3: ...an outlet on a circuit different from that to which the other device s is connected Consult the manufacturer at 1 800 638 3722 or 1 708 377 0600 or a field service technician for help DEFINITION OF SY...

Page 4: ...wer The LED next to the power jack will light up in red indicating that there is power to the scale Please refer to page 3 for information regarding the adapter available for use with this scale When...

Page 5: ...ht reading 1 Place the item you want to tare on the scale tray to weigh it 2 Press and release the ZERO CLEAR button to change the weight to 0 3 Place the baby on the scale with the tare item and weig...

Page 6: ...ocedure above OVER shows in display Overload The load on the scale exceeds scale maximum capacity 44 lbs 20 kg Remove weight turn scale off turn scale on again and try again UNDER shows in display Neg...

Page 7: ...mpering damage in transit or unauthorized repair or alternations Further the warranty does not cover Acts of God such as fire flood hurricanes and tornadoes This warranty gives you specific legal righ...

Page 8: ...zar NDICE DE MATERIAS ENGLISH 2 ESPA OL 8 FRAN AIS 14 ADVERTENCIAS Y PRECAUCIONES 9 S MBOLOS Y ESPECIFICACIONES 9 INFORMACI N DE LA CONECTIVIDAD 9 INSTRUCCI NES DE MONTAJE PILAS 10 INSTRUCCI NES B SIC...

Page 9: ...n un t cnico de servicio de campo o con la Asistencia T cnica de Health o meter Profesional al 1 800 638 3722 o 1 708 377 0600 DEFINICI N DE LOS S MBOLOS ESPECIFICACIONES DE LA BALANZA 553KL Capacidad...

Page 10: ...a fuente de alimentaci n La luz LED cerca al gato de la potencia se iluminar en rojo a indicar que la balanza est recibiendo alimentaci n del adaptador Refiera por favor a la p gina 9 para la informac...

Page 11: ...bjeto que quiere pesar como tara en la bandeja de la balanza a pesarlo 2 Apr te y suelte el bot n ZERO CLEAR para cambiar el peso a 0 3 Coloque el beb en la bandeja de la balanza con el objeto de tara...

Page 12: ...nza en una superficie nivelada 4 Compare el peso obtenido con el valor de peso conocido v a el procedimiento de calibraci n arriba La pantalla muestra OVER Demasiado El peso en la balanza excede la ca...

Page 13: ...cciones de uso maltratos inclusive alteraciones al equipo da os sufridos durante el transporte y reparaciones o modificaciones no autorizadas Adem s esta garant a tampoco cubre los da os causados por...

Page 14: ...ESPA OL 8 FRAN AIS 14 MISE EN GARDE ET AVERTISSEMENT 15 SYMBOLES ET SP CIFICATIONS D CETTE BALANCE 15 L INFORMATION DE CONNECTIVIT 15 CONSIGNES D INSTALLATION PILES 16 INSTRUCTIONS DE D MARRAGE RAPID...

Page 15: ...quipements Brancher l quipement une prise d un circuit lectrique qui est diff rent de celui auquel les autres appareils sont connect s Pour obtenir de l aide consulter le Support Technique Health o m...

Page 16: ...e La lumi re LED c t de la prise d alimentation s allume en rouge indiquant qu il a de puissance la balance Renvoyez s il vous pla t la page 15 pour l information en ce qui concerne les adaptateurs di...

Page 17: ...ter l interpr tation du poids du b b 1 Pesez l l ment que vous aimeriez tarer 2 Appuyer et rel chez le bouton ZERO CLEAR pour changer le poids 0 3 Placez le b b sur le plateau de la balance avec l l m...

Page 18: ...a charge exerc e sur la balance d passe sa capacit 20 kg 44 lbs Retirer la charge mettre la balance hors tension remettre la balance sous tension et essayer nouveau L cran affiche le message UNDER cha...

Page 19: ...avec une tension ou un courant inappropri s l utilisation contraire aux directives d utilisation l abus y compris l alt ration les dommages en transit ou les r parations ou les modifications non auto...

Page 20: ...LONG TERM CARE PEDIATRIC NEONATAL PRIMARY CARE PROFESSIONAL HOME HEALTH VETERINARY Please visit our web site to learn more about our full lines of professional scales Pelstar LLC 11800 South Austin A...

Reviews: