21
PME-MP70DP
A0658-7.3 en
HBM
4.5
Synchronization
Synchronization is recommended when
D
the transducer cables of several devices run side by side
D
the measuring points are unshielded and are close together
Synchronization prevents differences in the carrier frequency causing
disturbing superpositions.
1 2 3 4 5 6
ON
S10
1 2 34 5 6
ON
Master
Slave
S10
S10
Fig. 4.6:
Setting (master/slave) synchronization
When synchronizing several modules, one device (e.g. MP30,
not
the MP70)
must be declared the master. The other devices must be set as slaves. Even
if modules work without a CAN Bus, the flat ribbon cable should always be
used for synchronization between them.