Note
Calibration/Verification is valid with any calibrated Receiver channel.
Test results
Range Offset DCGain
SINL
MSE BWdth CMRR Noise ACCpl
(V)
(%)
(%)
(%)
(%)
(kHz)
(dB)
(%)
0.2
-
0.022
0.015
0.005 0.030
NA
NA
0.021
NA
0.4
-
0.008
0.030
0.004 0.023
NA
-
111.5
0.017
NA
1.0
-
0.003
0.005
0.004 0.006
NA
NA
0.015
NA
2.0
0.001
-
0.001
0.005 0.006
NA
NA
0.016 Passed
4.0
-
0.001
0.010
0.004 0.005 10181.9
-
107.0
0.018
NA
10.0
-
0.001
-
0.013
0.005 0.010
NA
NA
0.016
NA
20.0
0.001
-
0.016
0.005 0.009
NA
NA
0.016
NA
40.0
0.000
-
0.002
0.005 0.004
NA
-
113.1
0.018
NA
100.0
0.002
-
0.026
0.005 0.013
NA
NA
0.016
NA
Note
This table shows only a part of the report.
B.2.1
The Extra log file
Verification test report
When the Full
(extra)
test is selected, an
extra
log file will be present in the log
save folder with the
Full
bandwidth results inside.
ISOBE5600 Calibration and Verification Software: V2.10
Verification Results
Date
27
-
Apr
-
2010
Recorder Info
Physical Name
Recorder A
Serialnumber
IFA1234567
Type
ISOBE5600m Memory
SW version
2.00.10113
No. channels
4
ISOBE5600
I2716-4.0 en
111