B.3
Producing graphical results
-
DC Gain
The extra log file can be imported into excel to make a visual plot of the graphical
data.
Note
Be sure to select; Original data type = Delimited
►
Delimiter = Semi colon, to
be able to import the data correctly.
Figure B.1 is an example of the expected graphical output of the
DC Gain
section of a log file. (“Test name: DCgain test” on page 112).
Figure B.1:
Example of a graphical output (DC Gain Test)
One extra column (highlighted in blue) has been created to show the difference
between the two columns of the expected and actual results. Doing this will give
you the error in the results which you can add a trendline to (shown in red).
B.3.1
Producing graphical results
-
Full Bandwidth
When the Full
(extra)
test is selected, an
extra
log file will be present in the log
save folder with the
Full
bandwidth results inside.
This file can be imported into excel to make a visual plot of the graphical data.
ISOBE5600
I2716-4.0 en
115