Sultan Flow Series
MANUAL
Draft 1.2, Mar 2009
42
4
SECTION 5.7
MAXIMUM INTERVAL
This function sets a time limit for updating the totalised flow from an estimated value,
when the unit is in “CAL” or any other non measurement mode.
Non measurement mode: “NMM”
1. If NMM time is less than the Max Interval then the Totaliser will be updated with the
estimated value.
2. If NMM time is more than the Max Interval then the Totaliser will not be updated with
the estimated value.
The unit estimates the flow in the NMM period by averaging the flow values just prior to
and just after the end of period.
DISPLAY:
MAX INTERVAL
ENABLE
CAL
CAL
MAX TIME
12:34:56
CAL
MAX TIME
12
:34:56
0-24
CAL
MAX TIME
12:
34
:56
CAL
MAX TIME
12:34:
56
0-59
0-59
CLOCKSET
NO