24
Figure 5-11: Text Log Screen
5.3 Waveform Generator (cont’d)
5.3.4 Experiment Delivery Logs (cont’d)
5.3.4.1 Text Log
The text log contains the same information that appears on the
vaccine delivery process screen after completion. Text log data
is summarized in an easy-to-read format and is designed to
provide quick feedback regarding the delivery completion. An
example Log printout is given below in Figure 5-11.
Field
Description
Pulse
Pulse number
VSet
Voltage programmed to high voltage power supply
VMon
Voltage monitored during pulse
IMon
Current monitored during pulse
RMon
Equivalent load resistance calculated from voltage
and current monitors R = V / I
DurSet Pulse duration programmed to waveform generator
DurMon Pulse duration monitored by independent
time-auditor system
IntSet
Pulse interval programmed to waveform generator
IntMon Pulse interval monitored by independent
time-auditor system
Date Start: 8/14/2011
6:00:09 AM
(8/14/2011 1:00:09 PM GMT)
Experiment:
Pl
Transfectant: Plasmid A
Electrode:
IDA-2001
Operator:
BTX
Final Completion Code Code:
0 (Success)
++++++++++++++++++++++++++
Start:
8/14/2011
6:00:09 AM
(8/14/2011 1:00:09 PM GMT)
Process Completion Return Code: 0 (Success)
Pulse
VSet
PVMon
PIMon
RMon
DurSet
DurMon
IntSet
IntMon
1
450
450
0.351
1282
0.050
0.050
0.200
0.200
2
450
451
0.322
1401
0.050
0.050
50.000
49.856
3
110
110
0.141
783
10.000
9.975
20.000
19.944
4
110
109
0.146
750
10.000
9.975
20.000
19.944
5
110
108
0.147
736
10.000
9.975
20.000
19.944
6
110
108
0.141
767
10.000
9.975
20.000
19.944
7
110
108
0.132
821
10.000
9.975
20.000
19.944
8
110
108
0.131
827
10.000
9.975
20.000
19.944
9
110
108
0.159
681
10.000
9.975
20.000
19.947
10
110
109
0.144
756
10.000
9.975
20.000
0.000
Load Reading:
204
Date Stop: 2/12/2007 2:16:40 PM (2/12/2007 10:16:40 PM GMT)
Time Elapsed:
00:00:18
System Operation
5