26
5.3 Waveform Generator (cont’d)
5.3.5 Delivery Completion Return Codes
Return codes are available following system shutdown describing how it took place.
Return Code
Description
0
Successful completion (Normal Operation)
1
Emergency Flash memory storage failure
2
Waveform protocol data integrity verification error
3
Waveform protocol mode not recognized
4
Protocol voltage below minimum specification
5
Protocol voltage above maximum specification
6
Protocol pulse width below minimum specification
7
Protocol pulse width above maximum specification
8
Protocol pulse interval below minimum specification
9
Protocol pulse interval above maximum specification
10
Protocol pulse number below minimum specification
11
Protocol pulse number above maximum specification
12
Protocol has a zero pulse group in between two non-zero pulse groups
13
Protocol pulse settings exceed maximum duty cycle
14
Low voltage power supply failure (Hard system turn-off)
15
High volt power supply voltage above maximum voltage
16
High volt power supply maximum current draw
17
Pulse voltage monitor detected off-zero voltage while no pulse was active
18
Pulse voltage monitor below maximum droop specification
19
Pulse voltage monitor above maximum overshoot specification
20
Pulse current monitor below minimum specification
21
Pulse current monitor above maximum specification
22
Independent time-audit system failed to respond to query
23
Independent time-audit system detected less pulses than expected
24
Independent time-audit system detected more pulses than expected
25
Independent time-audit system exceeded maximum measurable pulse number
26
Independent time-audit system detected excessive pulse duration variance
27
Independent time-audit system detected excessive pulse interval variance
28
Independent time-audit system data integrity verification failure
29
Reset button on front panel triggered
30
Control system watchdog timeout
251
High voltage timeout
252
Load estimation timeout
253
Delivery process was terminated prematurely
254
Control system experienced a system reset of unknown source
255
Load estimate outside of allowable range
System Operation
5