5.3 Waveform Generator (cont’d)
5.3.4 Experiment Delivery Logs
All collected data from the therapeutic/vaccine delivery are automatically stored in the internal Flash Memory and
an inserted USB key under a unique chronological directory structure. Included items are an XML file containing raw
detailed system and runtime data, a text log with basic runtime data, and a comma-separated CSV file containing pulse
monitor data. The directory structure is constructed in a “AgilePulse\Log\<Year>\<Month>\<Day>\<Time>” hierarchy. A
maximum of 20,000 delivery logs can be stored in the internal system memory.
Raw Data
(CCEP <Date> <Time>.xml)
Text Log
(CCEP <Date> <Time>.log)
CSV Data
(CCEP <Date> <Time>.csv)
Start and Stop Date
•
•
Start and Stop Time
•
•
Experiment
•
•
•
Transfectent
•
•
•
Electrode ID
•
•
•
Process Completion Type
•
•
Set Pulse Parameters
•
•
•
Average monitor pulse voltage
•
Average monitor pulse current
•
Average monitor pulse resistance
•
All monitor pulse voltage samples
•
•
All monitor pulse current samples
•
•
Monitor width of each pulse
•
•
•
Monitor interval of each pulse
•
•
•
Pre-pulse Load Reading
•
•
System Serial Numbers
•
System Software Versions
•
System Operation
5