Installation 8

Typical installation examples with mounting accessories

Accessories are sold separately.

User-assembled, top thread immersion installation

I

II

1

2

4

3

5

6

To install the probe:

A.  Wrap Teflon™ tape around upper 

3/4

” NPT threads of the probe (6).

B.  Attach a user-supplied reducer bushing (5) to the probe (6) upper threads, 

step

 

I

C.  Feed probe cable through length of NPT externally threaded, user supplied pipe (4), matched to bushing.

D.  Thread the pipe into the upper threads of the reducing bushing (5) attached to the probe.

E.  The upper portion of the pipe should be sealed, 

step

 

II

, to prevent ingress of water if installed outdoor. 

F.  Attach the pipe to a handrail. 

Low thread immersion installation

The 

HI60501

 (PVC submersible holder) together with 

HI60501-2 

(protective end cap), and 

HI605011

 (mounting 

flange) provide a sturdy, protective housing designed for low-thread immersion installation. 

1 Cable gland
2 Pipe cap (socket connect or threaded)
3 Van Stone flange (one size smaller than the pipe)
4 2”, or similar, PVC pipe (schedule 80 PVC)

5

Reducer bushing 
•  internal thread to fit probe’s 

3/4

” NPT threads

•  external thread to fit pipe’s threads or 

socket-connect type

6 Optical dissolved oxygen probe 

Summary of Contents for HI7640-58

Page 1: ...nload the HI510 controller user manual https manuals hannainst com HI510 Hanna is committed to developing and deploying digital solutions with a positive impact on the environment and climate 0 to 50...

Page 2: ...essories 14 14 Abbreviations 14 Certification 15 Recommendations for Users 15 Warranty 15 Dear Customer Thank you for choosing a Hanna Instruments product Please read this instruction manual carefully...

Page 3: ...are all accessible inside controller s enclosure Before powering the controller verify the probe wiring has been done following the instructions detailed in this manual 3 GENERAL DESCRIPTION INTENDED...

Page 4: ...ed oxygen Over time the sensor s optical components can age but are compensated for by using the reference signal to compensate the measuring path As a result the sensor provides accurate dissolved ox...

Page 5: ...red by the user in saturation or mg L Temperature Range 5 0 to 50 0 C 23 0 to 122 0 F Resolution 0 1 C 0 1 F Accuracy 0 3 C 0 5 F Calibration Single point offset controller setting Temperature compens...

Page 6: ...Remove the protective cap 3 Locate the O ring that sits on probe body Sparingly lubricate with a thin film of supplied grease Use care to prevent grease fingerprints from contacting the optical window...

Page 7: ...stalled using the 3 4 NPT external threads Hand tighten the probe in position Then depending on the process tighten one or two turns with a wrench to secure in place Do not exceed 10 N m 7 3 lb ft tor...

Page 8: ...shing 5 attached to the probe E The upper portion of the pipe should be sealed step II to prevent ingress of water if installed outdoor F Attach the pipe to a handrail Low thread immersion installatio...

Page 9: ...temperature differences greater than 2 C 36 F between tank content and flow cell sample Shade the assembly from direct sunlight TANK VALVE 1 VALVE 2 FLOW FLOW INCOMING FLOW EXIT FLOW 3 1 2 4 5 7 6 6...

Page 10: ...de Marking Attached cable Patch cable Functionality GREEN BLACK 0V B WHITE WHITE RS485 D A YELLOW BLUE RS485 D BROWN RED 5V g GREEN YELLOW Protective ground connection See Accessories section for cabl...

Page 11: ...8 26 mg L suspend the probe in water saturated air This condition corresponds to 100 air saturated water at the temperature of measurement 1 Use of calibration beaker containing some water or moisten...

Page 12: ...lope if values are over 20 12 MAINTENANCE General Maintenance Inspect O ring for nicks or other damage Replacing the o ring is advised Do not substitute other grease or lubricants as it may cause the...

Page 13: ...e the new O ring from the container and slide on the probe tip do not roll or twist the O ring 8 Use a syringe filled with silicone grease and sparingly lubricate the O ring with a thin film of grease...

Page 14: ...10 m 32 9 HI76510 15 Patch cable 15 m 49 2 HI76510 25 Patch cable 25 m 82 HI76510 50 Patch cable 50 m 164 BL120 400 Flow cell probe adapter kit BL120 401 Flow cell valve BL120 402 Flow cell tubing 10...

Page 15: ...dous environments WARRANTY The probes are warrantied for two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions HI764113 1 S...

Page 16: ......

Reviews: