Hanna Instruments HI7640-58 Instruction Manual Download Page 11

11 Calibration

Standard Calibration

Preparation

•  Remove probe from process. 

•  Flush probe and cap with a jet of clean water.

•  Inspect for scratches or voids in cap surface.

•  Replace cap as required.

•  Shake any remaining solution off the probe. No droplets should remain on the DO sensing surface before 

performing the calibration procedure.

•  Verify the controller’s barometric pressure reading with a reference meter. Pressure value can be adjusted 

from the controller’s Technical Menu.

•  Verify the temperature is reading correctly. The temperature value can be adjusted from the controller’s 

Probe Setting menu.

•  Set Salinity value if exposing the probe to ocean or brackish waters.

Procedure

•  Set the controller to desired measurement unit.

Go to Probe Setting in Channel Setup and set required measurement mode (Meas.Mode): DO_Sat 
(% saturation) or DO_Conc (Concentration).

•  Press the 

key then 

CAL

 virtual key. The controller recognizes the currently selected measuring unit.

Calibration at 100 % and 0 % saturation

To calibrate at 100 % saturation (or 8.26 mg/L) suspend the probe in water-saturated air. This condition 
corresponds to 100 % air-saturated water at the temperature of measurement. 

1.  Use of calibration beaker containing some water or moistened absorbent material is recommended. 

Loosely screw the beaker onto the probe (first thread only).

2.  Allow 15 minutes for the air inside the beaker to become water saturated. 
3.  Press 

CAL

 to enter calibration mode after this 15 minute interval has elapsed.

The reading, temperature, atmospheric pressure (mmHg), calibration point and “Wait” message are displayed.

4.  Once the reading has stabilized the 

CFM

 virtual key is displayed. 

Press 

CFM

 to accept the calibration point. 

“Wait” followed by “Calibration point is accepted and saved” are displayed. 

Exit

 and 

Next

 virtual keys are available. Pressing 

Exit

 saves a single point calibration.

5.  Press 

Next

 to follow with second-point calibration.

6.  To calibrate at 0 % (or 0 mg/L), place the probe in the 

HI7040

 zero oxygen solution and stir gently for 

2-3 minutes. Dislodge bubbles that may adhere to the cap. 
“Wait” message and stability indicator (blinking) are displayed.

7.  Press 

CFM

 to accept the calibration point. 

“Calibration point is accepted and saved” message followed by “Calibration DONE” are displayed.

Summary of Contents for HI7640-58

Page 1: ...nload the HI510 controller user manual https manuals hannainst com HI510 Hanna is committed to developing and deploying digital solutions with a positive impact on the environment and climate 0 to 50...

Page 2: ...essories 14 14 Abbreviations 14 Certification 15 Recommendations for Users 15 Warranty 15 Dear Customer Thank you for choosing a Hanna Instruments product Please read this instruction manual carefully...

Page 3: ...are all accessible inside controller s enclosure Before powering the controller verify the probe wiring has been done following the instructions detailed in this manual 3 GENERAL DESCRIPTION INTENDED...

Page 4: ...ed oxygen Over time the sensor s optical components can age but are compensated for by using the reference signal to compensate the measuring path As a result the sensor provides accurate dissolved ox...

Page 5: ...red by the user in saturation or mg L Temperature Range 5 0 to 50 0 C 23 0 to 122 0 F Resolution 0 1 C 0 1 F Accuracy 0 3 C 0 5 F Calibration Single point offset controller setting Temperature compens...

Page 6: ...Remove the protective cap 3 Locate the O ring that sits on probe body Sparingly lubricate with a thin film of supplied grease Use care to prevent grease fingerprints from contacting the optical window...

Page 7: ...stalled using the 3 4 NPT external threads Hand tighten the probe in position Then depending on the process tighten one or two turns with a wrench to secure in place Do not exceed 10 N m 7 3 lb ft tor...

Page 8: ...shing 5 attached to the probe E The upper portion of the pipe should be sealed step II to prevent ingress of water if installed outdoor F Attach the pipe to a handrail Low thread immersion installatio...

Page 9: ...temperature differences greater than 2 C 36 F between tank content and flow cell sample Shade the assembly from direct sunlight TANK VALVE 1 VALVE 2 FLOW FLOW INCOMING FLOW EXIT FLOW 3 1 2 4 5 7 6 6...

Page 10: ...de Marking Attached cable Patch cable Functionality GREEN BLACK 0V B WHITE WHITE RS485 D A YELLOW BLUE RS485 D BROWN RED 5V g GREEN YELLOW Protective ground connection See Accessories section for cabl...

Page 11: ...8 26 mg L suspend the probe in water saturated air This condition corresponds to 100 air saturated water at the temperature of measurement 1 Use of calibration beaker containing some water or moisten...

Page 12: ...lope if values are over 20 12 MAINTENANCE General Maintenance Inspect O ring for nicks or other damage Replacing the o ring is advised Do not substitute other grease or lubricants as it may cause the...

Page 13: ...e the new O ring from the container and slide on the probe tip do not roll or twist the O ring 8 Use a syringe filled with silicone grease and sparingly lubricate the O ring with a thin film of grease...

Page 14: ...10 m 32 9 HI76510 15 Patch cable 15 m 49 2 HI76510 25 Patch cable 25 m 82 HI76510 50 Patch cable 50 m 164 BL120 400 Flow cell probe adapter kit BL120 401 Flow cell valve BL120 402 Flow cell tubing 10...

Page 15: ...dous environments WARRANTY The probes are warrantied for two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions HI764113 1 S...

Page 16: ......

Reviews: