2
1
3
DETAIL
1. Smart Cap
2. Alignment key
2. Sensor body
4. Temperature sensor
5. Embedded O
2
sensitive
luminophore with black
protective layer
6. RFID Tag
7. Optical window
8. O-Ring Seal
1 Smart Cap
2 Alignment key
3 Probe body
4 Temperature sensor
5 Embedded O
2
sensitive luminophore with rugged,
insoluble black oxygen-permeable protective layer
6 Data storage tag
7 Optical window
8 O-ring seal
Note
: Verify time and date are configured correctly in the controller’s Setup Menu prior to probe initialization.
9. INSTALLATION
General considerations
• Probes are easily installed using the
3/4
” NPT external threads.
• Hand tighten the probe in position. Then, depending on the process, tighten one or two turns with a
wrench to secure in place. Do not exceed 10 N
•
m (7.3 lb-ft) torque specification for the probe sensor.
• Protect the probe and the Smart Cap from strong flow to prevent unstable readings. In turbulent aeration
basin installations, place the probe in a weir for more accurate readings.
• Position the probe so that it does not trap air bubbles at the Smart Cap.
• Deposits of foreign material should not be allowed to accumulate within the sensing area.
• Protect the sensing surface from any blunt objects and keep it clean.
• Provisions must be made for the removal of the probe from the process
• Consider probe accessibility for maintenance when selecting placement.