Preliminary Examination

1.  PRELIMINARY EXAMINATION

Remove the probe and accessories from the packing and examine it carefully. 
Each probe is supplied with:

• 

HI764113-1

 DO Smart Cap with O-ring

•  Calibration beaker
•  Sachet with silicone grease
•  Syringe
•  Quick reference guide with probe quality certificate

Note

: Save all packing material until you are sure that the probe works correctly. Any damaged or defective item 

must be returned in its original packing material with the supplied accessories.

2.  SAFETY MEASURES

General Safety & Installation Recommendations

•  Electrical connection, installation, operation, and maintenance must be carried out by specialized 

personnel only. 
The specialized personnel must read, understand, and adhere to the instructions in this manual. 

•  Before connecting the probe to the process controller, disconnect the controller from the power line.
•  User serviceable connections are all accessible inside controller’s enclosure.
•  Before powering the controller, verify the probe wiring has been done following the instructions 

detailed in this manual.

3.  GENERAL DESCRIPTION & INTENDED USE

The 

HI7640-58

 series are optical dissolved oxygen probes with Smart Caps (

HI764113-1

) for measurement 

of dissolved oxygen.
The series is designed to work with the Hanna Instruments

®

 

HI510

 Universal Process Controller. When paired with 

the controller, the system provides accurate dissolved oxygen measurements auto-compensated for barometric 
pressure, salinity (set manually), and temperature.
Suitable for control applications in aeration basins, ponds, and tanks where optimizing oxygen transfer is a 
key element, the probe can be submersed/immersed in a tank using the 

3/4

” NPT threads, or installed in a 

flow cell using the lower sensor threads.
The probe is available with an 

integral connector

 that permits connections with no twisted cables as cables 

are connected after installation.

Patch cables

 may be purchased separately to connect between the probe and controller up to 50 meters (164 ft).

Summary of Contents for HI7640-58

Page 1: ...nload the HI510 controller user manual https manuals hannainst com HI510 Hanna is committed to developing and deploying digital solutions with a positive impact on the environment and climate 0 to 50...

Page 2: ...essories 14 14 Abbreviations 14 Certification 15 Recommendations for Users 15 Warranty 15 Dear Customer Thank you for choosing a Hanna Instruments product Please read this instruction manual carefully...

Page 3: ...are all accessible inside controller s enclosure Before powering the controller verify the probe wiring has been done following the instructions detailed in this manual 3 GENERAL DESCRIPTION INTENDED...

Page 4: ...ed oxygen Over time the sensor s optical components can age but are compensated for by using the reference signal to compensate the measuring path As a result the sensor provides accurate dissolved ox...

Page 5: ...red by the user in saturation or mg L Temperature Range 5 0 to 50 0 C 23 0 to 122 0 F Resolution 0 1 C 0 1 F Accuracy 0 3 C 0 5 F Calibration Single point offset controller setting Temperature compens...

Page 6: ...Remove the protective cap 3 Locate the O ring that sits on probe body Sparingly lubricate with a thin film of supplied grease Use care to prevent grease fingerprints from contacting the optical window...

Page 7: ...stalled using the 3 4 NPT external threads Hand tighten the probe in position Then depending on the process tighten one or two turns with a wrench to secure in place Do not exceed 10 N m 7 3 lb ft tor...

Page 8: ...shing 5 attached to the probe E The upper portion of the pipe should be sealed step II to prevent ingress of water if installed outdoor F Attach the pipe to a handrail Low thread immersion installatio...

Page 9: ...temperature differences greater than 2 C 36 F between tank content and flow cell sample Shade the assembly from direct sunlight TANK VALVE 1 VALVE 2 FLOW FLOW INCOMING FLOW EXIT FLOW 3 1 2 4 5 7 6 6...

Page 10: ...de Marking Attached cable Patch cable Functionality GREEN BLACK 0V B WHITE WHITE RS485 D A YELLOW BLUE RS485 D BROWN RED 5V g GREEN YELLOW Protective ground connection See Accessories section for cabl...

Page 11: ...8 26 mg L suspend the probe in water saturated air This condition corresponds to 100 air saturated water at the temperature of measurement 1 Use of calibration beaker containing some water or moisten...

Page 12: ...lope if values are over 20 12 MAINTENANCE General Maintenance Inspect O ring for nicks or other damage Replacing the o ring is advised Do not substitute other grease or lubricants as it may cause the...

Page 13: ...e the new O ring from the container and slide on the probe tip do not roll or twist the O ring 8 Use a syringe filled with silicone grease and sparingly lubricate the O ring with a thin film of grease...

Page 14: ...10 m 32 9 HI76510 15 Patch cable 15 m 49 2 HI76510 25 Patch cable 25 m 82 HI76510 50 Patch cable 50 m 164 BL120 400 Flow cell probe adapter kit BL120 401 Flow cell valve BL120 402 Flow cell tubing 10...

Page 15: ...dous environments WARRANTY The probes are warrantied for two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions HI764113 1 S...