Hanna Instruments HI7640-58 Instruction Manual Download Page 5

5 

Specifications

6.  SPECIFICATIONS

Range

0.00 to 50.00 mg/L (ppm) concentration

0.0 to 500.0 % saturation

Resolution

0.01 mg/L (ppm) 

0.1 % saturation

Accuracy

from 0.00 to 20.00 mg/L (ppm)  1.5 % of reading or ± 0.01 mg/L (ppm), whichever is greater

from 20.00 to 50.00 mg/L (ppm)  ±5% of reading
from 0.0 to 200.0 % saturation   ± 1.5 % of reading or ± 0.1 %, whichever is greater

from 200.0 to 500.0 % saturation  ±5 % of reading

Calibration

One or two points 

at 100 % and/or 0 % or 8.26 mg/L and/or 0 mg/L  

 

in water-saturated air or zero-oxygen solution

Single point process calibration 

using value entered by the user in % saturation or mg/L

Temperature

Range 

−5.0 to 50.0 °C (23.0 to 122.0 °F)

Resolution 

0.1 °C / 0.1 °F

Accuracy 

±0.3 °C / ±0.5 °F

Calibration 

Single point offset (controller setting)

Temperature 

compensation Automatic

Temperature 

source

Automatic (from probe)

Manual

Pressure 

compensation Automatic 

420 to 850 mmHg

Salinity 

compensation Automatic 

0 to 70 g/L  (salinity factor set by user)

Body

ABS

Sensor

Type 

Optical DO

Diameter 

Ø 17 mm (0.7”)

Insertion length 

43 mm (1.7”)

Maximum 

pressure

2 bar (29 psi) at 25 °C (77 °F)

Threaded 

connection

3/4

” NPT external thread (both ends)

Wetted parts

Sensor body 

PVDF

Smart cap material 

Polypropylene

Dome-shaped membrane 

PMMA

O-ring NBR

Temperature contact 

Stainless steel 

Protection 

rating

IP68

Cable length See 

zz

, Series Configuration

Summary of Contents for HI7640-58

Page 1: ...nload the HI510 controller user manual https manuals hannainst com HI510 Hanna is committed to developing and deploying digital solutions with a positive impact on the environment and climate 0 to 50...

Page 2: ...essories 14 14 Abbreviations 14 Certification 15 Recommendations for Users 15 Warranty 15 Dear Customer Thank you for choosing a Hanna Instruments product Please read this instruction manual carefully...

Page 3: ...are all accessible inside controller s enclosure Before powering the controller verify the probe wiring has been done following the instructions detailed in this manual 3 GENERAL DESCRIPTION INTENDED...

Page 4: ...ed oxygen Over time the sensor s optical components can age but are compensated for by using the reference signal to compensate the measuring path As a result the sensor provides accurate dissolved ox...

Page 5: ...red by the user in saturation or mg L Temperature Range 5 0 to 50 0 C 23 0 to 122 0 F Resolution 0 1 C 0 1 F Accuracy 0 3 C 0 5 F Calibration Single point offset controller setting Temperature compens...

Page 6: ...Remove the protective cap 3 Locate the O ring that sits on probe body Sparingly lubricate with a thin film of supplied grease Use care to prevent grease fingerprints from contacting the optical window...

Page 7: ...stalled using the 3 4 NPT external threads Hand tighten the probe in position Then depending on the process tighten one or two turns with a wrench to secure in place Do not exceed 10 N m 7 3 lb ft tor...

Page 8: ...shing 5 attached to the probe E The upper portion of the pipe should be sealed step II to prevent ingress of water if installed outdoor F Attach the pipe to a handrail Low thread immersion installatio...

Page 9: ...temperature differences greater than 2 C 36 F between tank content and flow cell sample Shade the assembly from direct sunlight TANK VALVE 1 VALVE 2 FLOW FLOW INCOMING FLOW EXIT FLOW 3 1 2 4 5 7 6 6...

Page 10: ...de Marking Attached cable Patch cable Functionality GREEN BLACK 0V B WHITE WHITE RS485 D A YELLOW BLUE RS485 D BROWN RED 5V g GREEN YELLOW Protective ground connection See Accessories section for cabl...

Page 11: ...8 26 mg L suspend the probe in water saturated air This condition corresponds to 100 air saturated water at the temperature of measurement 1 Use of calibration beaker containing some water or moisten...

Page 12: ...lope if values are over 20 12 MAINTENANCE General Maintenance Inspect O ring for nicks or other damage Replacing the o ring is advised Do not substitute other grease or lubricants as it may cause the...

Page 13: ...e the new O ring from the container and slide on the probe tip do not roll or twist the O ring 8 Use a syringe filled with silicone grease and sparingly lubricate the O ring with a thin film of grease...

Page 14: ...10 m 32 9 HI76510 15 Patch cable 15 m 49 2 HI76510 25 Patch cable 25 m 82 HI76510 50 Patch cable 50 m 164 BL120 400 Flow cell probe adapter kit BL120 401 Flow cell valve BL120 402 Flow cell tubing 10...

Page 15: ...dous environments WARRANTY The probes are warrantied for two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions HI764113 1 S...