Hanna Instruments HI7640-58 Instruction Manual Download Page 6

Probe Dimensions & Cable Connection 

6

7.  PROBE DIMENSIONS & CABLE CONNECTION

3/4’’ NPT

Ø 17 mm (0.7”)

Ø 28 mm (1.1”)

3/4’’ NPT

172 mm (6.7”)

64 mm (2.5”)

43 mm (1.7”)

172 mm (6.7”)

Align the pins and key then push the plug into the socket. 
Rotate the collar to lock in place. 

Note

: Probe connection (probe with integral DIN connector) and probe wiring (probe with attached cable) 

must be

 

carried out with the controller 

disconnected

 from power.

8.  PROBE PREPARATION & CONDITIONING

Note

: Read all the steps prior to starting probe preparation.

1.  Invert the probe so the cable faces the floor. 
2.  Remove the protective cap. 
3.  Locate the O-ring that sits on probe body. 

Sparingly lubricate with a thin film of supplied grease. Use care to prevent grease/fingerprints from 
contacting the optical window.

4.  Remove the Smart Cap from the container.
5.  Align the notched cutout arrow on the Smart Cap with the matching guide on the probe body.
6.  Slide and press the Smart Cap onto the probe’s body until the cap snaps in place. 

Once the cap is installed, it should not be removed unless a new cap is required.

7.  Place the probe in purified water for a minimum of 2 hours to hydrate the Smart Cap before use.

Note

: If the probe will not be installed immediately, place in the calibration/storage vessel with fresh water to 

protect it from damage and hydrate the cap. 

Summary of Contents for HI7640-58

Page 1: ...nload the HI510 controller user manual https manuals hannainst com HI510 Hanna is committed to developing and deploying digital solutions with a positive impact on the environment and climate 0 to 50...

Page 2: ...essories 14 14 Abbreviations 14 Certification 15 Recommendations for Users 15 Warranty 15 Dear Customer Thank you for choosing a Hanna Instruments product Please read this instruction manual carefully...

Page 3: ...are all accessible inside controller s enclosure Before powering the controller verify the probe wiring has been done following the instructions detailed in this manual 3 GENERAL DESCRIPTION INTENDED...

Page 4: ...ed oxygen Over time the sensor s optical components can age but are compensated for by using the reference signal to compensate the measuring path As a result the sensor provides accurate dissolved ox...

Page 5: ...red by the user in saturation or mg L Temperature Range 5 0 to 50 0 C 23 0 to 122 0 F Resolution 0 1 C 0 1 F Accuracy 0 3 C 0 5 F Calibration Single point offset controller setting Temperature compens...

Page 6: ...Remove the protective cap 3 Locate the O ring that sits on probe body Sparingly lubricate with a thin film of supplied grease Use care to prevent grease fingerprints from contacting the optical window...

Page 7: ...stalled using the 3 4 NPT external threads Hand tighten the probe in position Then depending on the process tighten one or two turns with a wrench to secure in place Do not exceed 10 N m 7 3 lb ft tor...

Page 8: ...shing 5 attached to the probe E The upper portion of the pipe should be sealed step II to prevent ingress of water if installed outdoor F Attach the pipe to a handrail Low thread immersion installatio...

Page 9: ...temperature differences greater than 2 C 36 F between tank content and flow cell sample Shade the assembly from direct sunlight TANK VALVE 1 VALVE 2 FLOW FLOW INCOMING FLOW EXIT FLOW 3 1 2 4 5 7 6 6...

Page 10: ...de Marking Attached cable Patch cable Functionality GREEN BLACK 0V B WHITE WHITE RS485 D A YELLOW BLUE RS485 D BROWN RED 5V g GREEN YELLOW Protective ground connection See Accessories section for cabl...

Page 11: ...8 26 mg L suspend the probe in water saturated air This condition corresponds to 100 air saturated water at the temperature of measurement 1 Use of calibration beaker containing some water or moisten...

Page 12: ...lope if values are over 20 12 MAINTENANCE General Maintenance Inspect O ring for nicks or other damage Replacing the o ring is advised Do not substitute other grease or lubricants as it may cause the...

Page 13: ...e the new O ring from the container and slide on the probe tip do not roll or twist the O ring 8 Use a syringe filled with silicone grease and sparingly lubricate the O ring with a thin film of grease...

Page 14: ...10 m 32 9 HI76510 15 Patch cable 15 m 49 2 HI76510 25 Patch cable 25 m 82 HI76510 50 Patch cable 50 m 164 BL120 400 Flow cell probe adapter kit BL120 401 Flow cell valve BL120 402 Flow cell tubing 10...

Page 15: ...dous environments WARRANTY The probes are warrantied for two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions HI764113 1 S...

Page 16: ......

Reviews: