19

18

the value will be stored in the lot 00 (log on demand data area). If
the print function is enabled, a printout is also produced providing the
following information:

a - Date
b - Sample number
c - Time
d - Readings ("----" means out of

range or probe not connected)

Note:

When the Log on demand data area is full and the LOG key is
pressed, the sample will not be stored and the LCD will display
"FULL". In this case it is necessary to delete the Log on demand
data to proceed.

TO VIEW LOGGING INFORMATION

If the ALT and LOT INFO keys are pressed during logging, the meter
displays for a few seconds the current lot and the number of logged
samples. The meter then returns to normal operational mode auto-
matically.

If ALT and LOT INFO are pressed while the meter is not logging, the
last logged lot in the lower LCD and the number of logged samples in
the upper LCD are displayed. It is then possible to scroll through the
following lot information with the 

"

 key:

• lot starting date (dd.mm)
• lot starting year
• lot starting time (hh.mm)
• lot logging interval
• logged channels (HI 98804 only)
By pressing the 

"

 key again, the meter displays the number of

logged samples again.
When the number of logged samples is displayed, it possible to pass
to another lot with the 

#

 and 

!

 keys. Press 

!

 to view the older

lots or 

#

 to view the more recent ones.

If 

#

 is pressed when the last lot is displayed, the meter displays the

lot 00 (log on demand). By pressing the 

#

 key again, the meter will

pass to the oldest lot.

Note

For lot 00, only the number of samples will be displayed.

To exit from the logging info viewing mode press ALT and LOT INFO
again or CH/TIME (HI98804 only) or TEMP/TIME (HI98801 only).

logged reading will appear briefly on the LCD. To reactivate the
LCD press ON/OFF.

TO STOP LOGGING

In order to stop the recording mode, press
ALT and LOG  keys (press ON/OFF first, if
meter is in sleep mode).
A last printout reporting the number of logged
samples (e.g. S.00009) will be produced if
printer is enabled.

Notes:

• It is recommended to use the adapter during logging in printing

mode, especially when many printouts are going to be taken.

• Before proceeding with logging with printing, make sure there is

enough paper for your measurements. When the paper is finished
the meter will not advise the operator and the printouts could be
lost. If this happens, data will continue to be stored in memory,
and it is always possible to print the data at a later time through
setup code 00.

• It is possible to insert a new paper roll during logging session.
• Once in the logging mode, the interval cannot be changed. Exit

the logging mode first (pressing the ALT and LOG keys) to set a
new interval.

• If the LOG key is pressed while in logging with printing mode, a

printout is produced without affecting the running sample number
and the value is stored in Log on demand area.

LOW BATTERY CONDITION

Printout is automatically disabled when battery charge weakens. The
last message "Stop log" will be printed and data will continue to be
stored in memory with the LOG and LOBAT
symbols blinking on LCD. If the user attempts
to enable the printer while in low battery
condition the message "bAtt" will appear for
a few seconds on the LCD.

Note:

When an external adapter or new batteries are connected, the
printing must be manually enabled in order to return to logging
with printing mode.

LOG ON DEMAND

In measuring or Timed log mode, press LOG to store
the current reading. The LCD will display "Stor" and

Summary of Contents for HI 98701

Page 1: ...nual HI 98701 HI 98801 HI 98704 HI 98804 Portable Microprocessor Printing and Logging K J T Thermocouple Thermometers h t t p w w w h a n n a i n s t c o m These Instruments are in Compliance with the...

Page 2: ...may be required The meters provide a controlled access to calibration and GLP settings through a password protection method PRELIMINARY EXAMINATION PRELIMINARY EXAMINATION PRELIMINARY EXAMINATION PREL...

Page 3: ...aker To prolong battery life the backlight and printing features are dis abled when the batteries are getting low LOBAT indication is displayed on LCD to warn the user of this condition However the me...

Page 4: ...tivate the display press the ON OFF key Note When the use of an alternate function FNC CFM and LOT INFO is requested press and hold the ALT key first and then the second listed key ON OFF SPECIFICATIO...

Page 5: ...heir valid values and the factory settings default Code Valid values Default 00 Lot data printing scrolling1 00 16 00 01 Print lots data summary1 02 Printer enable1 On enabled Off disabled On 03 Loggi...

Page 6: ...l the printer stops During printout the LCD will display the sample number being printed at that moment If printout is stopped the LCD will show Code Valid values Default 10 Show GLP data 11 Calibrati...

Page 7: ...t code 70 The types are indicated with dEF for K type J for J type and t for T type Choose the desired type using the arrow keys Press ALT and CFM to confirm the value the last printed sample number I...

Page 8: ...nel only for technical personnel only for technical personnel only for technical personnel only for technical personnel only To prepare the instrument for use choose the most appropriate tem perature...

Page 9: ...ING FUNCTIONS PRINTING LOGGING FUNCTIONS PRINTING LOGGING FUNCTIONS THERMOCOUPLE CALIBRATION 1 With a mV simulator input 0 000 mV at channel 1 of the meter 2 Ensure the meter is not displaying the col...

Page 10: ...04 only or TEMP TIME HI98801 only logged reading will appear briefly on the LCD To reactivate the LCD press ON OFF TO STOP LOGGING In order to stop the recording mode press ALT and LOG keys press ON O...

Page 11: ...print on demand to 001 see TO RESET PRINTING SAMPLE NUMBER in SETUP MODE paragraph HI 98701 HI 98704 HI 98701 HI 98704 HI 98701 HI 98704 HI 98701 HI 98704 HI 98701 HI 98704 PRINTING FUNCTIONS PRINTIN...

Page 12: ...98804 only only only only only OTHER FEATURES OTHER FEATURES OTHER FEATURES OTHER FEATURES OTHER FEATURES LCD BACKLIGHT The LCD can be illuminated to allow the user to see the readings even in dark en...

Page 13: ...case the starting time date and the dimension of the old lot are updated Note The oldest lot data are erased without any warning message Note Timed logging memory can be entirely erased through the s...

Page 14: ...g the batteries or connecting the external power adapter the meter will show batt on LCD Battery replacement must only take place in a non hazardous area using 1 5V alkaline AA type batteries In order...

Page 15: ...0 F 420mm 13 3 280mm 11 100mm 4 16 5 400mm 15 7 260mm 10 2 155mm 6 1 3mm 0 12 200mm 7 8 5mm 0 20 300mm 11 8 30mm 1 2 130mm 5 1 8mm 0 3 HI 766B1 90 surface probe max 450 C 840 F HI 766B2 Surface probe...

Page 16: ...ed of the charges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization number from the Customer Service department and then send it with shippi...

Page 17: ...w w w h a n n a i n s t c o m PRINTED IN PORTUGAL HANNA LITERATURE HANNA LITERATURE HANNA LITERATURE HANNA LITERATURE HANNA LITERATURE LAB RECORDING WATER ANALYSIS HANDBOOK WATER TREATMENT GENERAL CA...

Reviews: