background image

11

10

cod. 33: Year

cod. 40: Auto OFF

cod. 41: Battery test

cod. 50: Baud rate

cod. 60: Release code

cod. 70: Thermocouple type

cod. 71: Celsius or Fahrenheit

cod. 99: Pass Code

Some of the most important functions are explained below in a step by
step sequence.

TO SCAN LOGGED DATA

 (HI98801 and HI98804 only)

COD. 00 - Lot data Printing / Scrolling

• Select the code 00.
• The message "Lot data Printing" will scroll twice across LCD.

• The upper LCD will then display L 00

with the 00 blinking.

• Set the desired lot by the arrow keys. L00 is the lot of data of the

"log on demand" and L 01 to L 16 are the lots of the "timed log".

• Press the ALT and CFM keys to confirm the lot number.
• If the lot doesn't contain data, the "no data" message will scroll

across the LCD twice and the meter will return to setup mode.

• If the lot contains one or more data the

LCD will display the sample number in
its upper part and Sn in the lower part.

Note:

In the L 00 lot (log on demand) the sample number will be
displayed with 3 digits (001).

• Select the sample number to scan by the arrow keys.

Printing logged data

• Press ALT and CFM to print logged data.
• If the selected sample number is invalid (equal to 0 or bigger than

the number of samples), the "Err" message will be displayed for a
few seconds.

• If the sample number is correct, the samples starting from the

selected one to the last sample of the lot will be printed. To stop
printing before the last sample is reached, press and hold the ALT
and PAPER keys until the printer stops.

• During printout the LCD will display the sample number being

printed at that moment. If printout is stopped the LCD will show

Code

Valid values

Default

10 Show GLP data
11 Calibration alarm time-out

On(enabled); Off(disabled) On

20 Instrument ID code

0000÷9999

0000

30 Current time

3

hh:mm

00:00

31 Current day

3

dd

01

32 Current month

3

MM

01

33 Current year

3

YYYY

1998

40 Auto-Off/Power down time-out Off,5,10,15,30,45,60

5

41 Battery level test
50 RS232 baud rate

1

1200, 2400, 4800

4800

60 Firmware version
70 Thermocouple type selection

"dEF"(K) ; "J"(J) ; "t"(T) "dEF"

71 Celsius/Fahrenheit selection

°C ; °F

°C

99 Password

4

0000÷9999

0000

Note:

If a wrong code is entered the "Err" message is displayed on
LCD for a few seconds.

Note:

Code 05 has a different function in HI9880x and HI9870x
series as explained in detail in the following.

1

In HI 98801 and HI 98804 only.

2

In HI 98701 and HI 98704 only.

3

The meter automatically checks for entered time/date accuracy as follows:

      0

hh

23; 0

mm

59; 01

dd

28/29/30/31; 1

MM

12; 1998

YYYY

2097.

4

To change the password, the correct code must be entered first. If the password

has been forgotten, the password protected features are no longer accessible; in

this case contact your nearest Hanna Service Center.

SETUP MESSAGES LIST

cod. 00: Lot data Printing

cod. 01: Lot table Printing

cod. 02: Printer enable

cod. 03: Log Interval

cod. 05: Press "ALT CFM" to reset the sample number or "ALT FNC" to escape

2

       or: Press "ALT CFM" to delete Lot00 or "ALT FNC" to escape

1

cod. 06: Press "ALT CFM" to delete Lot  01-16 or "ALT FNC" to escape

cod. 10: GLP

cod. 11: Calibration alarm time-out

cod. 20: Instrument ID Code

cod. 30: Hour - Minute

cod. 31: Day

cod. 32: Month

Summary of Contents for HI 98701

Page 1: ...nual HI 98701 HI 98801 HI 98704 HI 98804 Portable Microprocessor Printing and Logging K J T Thermocouple Thermometers h t t p w w w h a n n a i n s t c o m These Instruments are in Compliance with the...

Page 2: ...may be required The meters provide a controlled access to calibration and GLP settings through a password protection method PRELIMINARY EXAMINATION PRELIMINARY EXAMINATION PRELIMINARY EXAMINATION PREL...

Page 3: ...aker To prolong battery life the backlight and printing features are dis abled when the batteries are getting low LOBAT indication is displayed on LCD to warn the user of this condition However the me...

Page 4: ...tivate the display press the ON OFF key Note When the use of an alternate function FNC CFM and LOT INFO is requested press and hold the ALT key first and then the second listed key ON OFF SPECIFICATIO...

Page 5: ...heir valid values and the factory settings default Code Valid values Default 00 Lot data printing scrolling1 00 16 00 01 Print lots data summary1 02 Printer enable1 On enabled Off disabled On 03 Loggi...

Page 6: ...l the printer stops During printout the LCD will display the sample number being printed at that moment If printout is stopped the LCD will show Code Valid values Default 10 Show GLP data 11 Calibrati...

Page 7: ...t code 70 The types are indicated with dEF for K type J for J type and t for T type Choose the desired type using the arrow keys Press ALT and CFM to confirm the value the last printed sample number I...

Page 8: ...nel only for technical personnel only for technical personnel only for technical personnel only for technical personnel only To prepare the instrument for use choose the most appropriate tem perature...

Page 9: ...ING FUNCTIONS PRINTING LOGGING FUNCTIONS PRINTING LOGGING FUNCTIONS THERMOCOUPLE CALIBRATION 1 With a mV simulator input 0 000 mV at channel 1 of the meter 2 Ensure the meter is not displaying the col...

Page 10: ...04 only or TEMP TIME HI98801 only logged reading will appear briefly on the LCD To reactivate the LCD press ON OFF TO STOP LOGGING In order to stop the recording mode press ALT and LOG keys press ON O...

Page 11: ...print on demand to 001 see TO RESET PRINTING SAMPLE NUMBER in SETUP MODE paragraph HI 98701 HI 98704 HI 98701 HI 98704 HI 98701 HI 98704 HI 98701 HI 98704 HI 98701 HI 98704 PRINTING FUNCTIONS PRINTIN...

Page 12: ...98804 only only only only only OTHER FEATURES OTHER FEATURES OTHER FEATURES OTHER FEATURES OTHER FEATURES LCD BACKLIGHT The LCD can be illuminated to allow the user to see the readings even in dark en...

Page 13: ...case the starting time date and the dimension of the old lot are updated Note The oldest lot data are erased without any warning message Note Timed logging memory can be entirely erased through the s...

Page 14: ...g the batteries or connecting the external power adapter the meter will show batt on LCD Battery replacement must only take place in a non hazardous area using 1 5V alkaline AA type batteries In order...

Page 15: ...0 F 420mm 13 3 280mm 11 100mm 4 16 5 400mm 15 7 260mm 10 2 155mm 6 1 3mm 0 12 200mm 7 8 5mm 0 20 300mm 11 8 30mm 1 2 130mm 5 1 8mm 0 3 HI 766B1 90 surface probe max 450 C 840 F HI 766B2 Surface probe...

Page 16: ...ed of the charges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization number from the Customer Service department and then send it with shippi...

Page 17: ...w w w h a n n a i n s t c o m PRINTED IN PORTUGAL HANNA LITERATURE HANNA LITERATURE HANNA LITERATURE HANNA LITERATURE HANNA LITERATURE LAB RECORDING WATER ANALYSIS HANDBOOK WATER TREATMENT GENERAL CA...

Reviews: