background image

Instruction Manual

  &  

(HI 983302)                (HI 983301)

On-line, Waterproof

EC-TDS meter with Alarm

http://www.hannainst.com

These Instruments are in

Compliance with the CE Directives

Hanna Instruments reserves the right to modify the design, construction

and appearance of its products without advance notice.

WARRANTY

WARRANTY

WARRANTY

WARRANTY

WARRANTY

EC and TDS  

  are warranted for two years 

against

defects in workmanship and materials when used for their intended

purpose and maintained according to instructions.

 The probes are

warranted for a period of one year.

 This warranty is limited to

repair or replacement free of charge.

Damages due to accident, misuse, tampering or lack of prescribed

maintenance are not covered.

If service is required, contact the dealer from whom you purchased the

instrument. If under warranty, report the model number, date of pur-

chase, serial number and the nature of the failure. If the repair is not

covered by the warranty, you will be notified of the charges incurred. If

the instrument is to be returned to Hanna Instruments, first obtain a

Returned Goods Authorization Number from the Customer Service de-

partment and then send it with shipment costs prepaid. When shipping

any instrument, make sure it is properly packaged for complete protec-

tion.
To validate your warranty, fill out and return the enclosed warranty card

within 14 days from the date of purchase.

All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instru-
ments Inc., Woonsocket, Rhode Island, 02895 , USA.

Dear Customer,
Thank you for choosing a Hanna product. This manual will

provide you with the necessary information for a correct

operation. Please read it carefully before using the meter. If

you need additional technical information, do not hesitate

to e-mail us at [email protected].
These instruments are in compliance with the   directives

EN 50081-1 and EN 50082-1.

Remove the instrument from the packing material and

examine it carefully. If any damage has occurred during

shipment, immediately notify your Dealer or the nearest

Hanna Customer Service Center.
The meters are supplied with:

• HI 7632 EC probe (for HI 983302);

• HI 7634 TDS probe (for HI 983301);

• Calibration screwdriver;

• 12 VDC power adapter.

Note:

Conserve all packing material until the instrument

has been observed to function correctly. Any defective

item must be returned in its original packing.

PRELIMINARY EXAMINATION

PRELIMINARY EXAMINATION

PRELIMINARY EXAMINATION

PRELIMINARY EXAMINATION

PRELIMINARY EXAMINATION

EC and TDS 

 

are instruments specially de-

signed to meet the needs of growers in greenhouses and

hydroponic applications.
The housing has been completely sealed against vapors and

humidity with IP54 rating.
You can simply hang the meter right above the sample to be

tested for continuous measurement.
Both meters come with a probe that compensates for the

temperature variation automatically. The probe is easy to

clean and requires little maintenance.
You can even select your own setpoint and be alerted of an

abnormal situation with a flashing LED alarm.
Measurements are highly accurate and the meters can be

calibrated at one point. In addition, the 

TDS  

uses a specially devised formula which perfectly reflects the

needs of growers. In fact, the common method of measuring

Total Dissolved Solids by multiplying the conductivity by 0.5

is far off the mark in horticulture with high concentrations of

phosphate, nitrogen, etc.
You no longer need to worry about battery changes either:

the unit runs without interruption on 12 VDC power supply.

GENERAL DESCRIPTION

GENERAL DESCRIPTION

GENERAL DESCRIPTION

GENERAL DESCRIPTION

GENERAL DESCRIPTION

FUNCTIONAL DESCRIPTION

FUNCTIONAL DESCRIPTION

FUNCTIONAL DESCRIPTION

FUNCTIONAL DESCRIPTION

FUNCTIONAL DESCRIPTION

1. Molded eye
2. Alarm LED

3. EC or TDS probe
4. 12 VDC power adapter

5. Power supply connector
6. Calibration trimmer

7. Liquid Crystal Display

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

EC 

 – HI 983302

Range

0.00 to 9.99 mS/cm (EC)

Resolution

0.01 mS/cm (EC)

Accuracy (@ 25°C/77°F)

±2% f.s.

Typical EMC Deviation

±2% f.s.

Temperature Compensation

Automatic

5 to 50°C (41 to 122°F)

Setpoint

0.70 to 3.50 mS/cm (EC)

Hysteresis

±0.20 mS/cm (EC) around setpoint

Alarm

LED blinks when EC value is outside

hysteresis range

Calibration

Manual with one trimmer

Probe

HI 7632 EC probe (included)

Casing

IP54

Power supply

External 12 VDC (included)

Dimensions

86 x 94 x 33 mm (3.4 x 3.7 x 1.3”)

Weight

150 g (5.3 oz.)

TDS 

 – HI 983301

Range

0 to 1990 mg/L (ppm)

Resolution

10 mg/L (ppm)

Accuracy (@ 25°C/77°F)

±2% f.s.

Typical EMC Deviation

±2% f.s.

TDS Factor

0.7

Temperature Compensation

Automatic

5 to 50°C (41 to 122°F)

Setpoint

500 to 1600 mg/L (ppm)

Hysteresis

±100 ppm around setpoint

Alarm

LED blinks when TDS value is outside

hysteresis range

Calibration

Manual with one trimmer

Probe

HI 7634 TDS probe (included)

Casing

IP54

Power supply

External 12 VDC (included)

Dimensions

86 x 94 x 33 mm (3.4 x 3.7 x 1.3”)

Weight

150 g (5.3 oz.)