TAKING MEASUREMENTS

• Turn the meter on by connecting the 12 VDC power

adapter to the meter and to the mains.

• Immerse the tip (4 cm/1½”)

of the EC or TDS probe in the
sample.

Note:

 In order not to affect the

accuracy of measurements,

it is important that the
probe body does not touch
nor stand close to the side
walls of the vessel. The tip
may lay on the bottom of
the beaker.

• The LCD will show the EC or

TDS value. Any initial varia-
tion may be due to temperature
compensation and the fact that
the probe is adjusting itself to
the new sample. Allow the read-
ing to stabilize and the meter

will start continuous monitor-
ing.

ADJUSTING THE SETPOINT

With the 

EC 

and 

TDS  

, you can select your own

setpoint and be alerted with a visual LED alarm when an
abnormal situation arises.
• Unscrew and remove the rear panel and gasket seal to

access the MEASURE/SET switch.

• Move the switch to the left (SET Mode).

• With a small screwdriver adjust the setpoint trimmer to

display the desired value in the 0.70 to 3.50 mS/cm or
in the 500 to 1600 ppm range for the 

EC 

and 

TDS

,

 

respectively.

• Make sure the switch is moved back to

the right (MEASURE Mode).

• Replace the rear panel and

the gasket, ensuring the unit

is properly closed.

• Whenever the EC or TDS read-

ing varies by more than

±0.20 mS/cm or ±100 ppm
from the setpoint, respectively,
the red alarm LED will blink.

PROBE MAINTENANCE

To minimize clogging and provide longer life for the probe,
it is recommended to clean it often or at least once a month.
• Immerse the tip of the electrode in 

HI 7061

 Cleaning

Solution for one hour or clean it with detergents used to

wash irrigation pipelines and fertilizer vessels.

• If a more thorough cleaning is required, brush the

metal pins with very fine sandpaper.

• After cleaning or when not in use, rinse the probe with

tap water.

For the greatest accuracy, frequent calibration of the instru-
ment is recommended.
• Turn the meter on and make sure that

the MEASURE/SET switch is on the MEA-
SURE mode.

• Pour a small quantity of

1413 µS/cm calibration so-
lution or 1500 ppm TDS
calibration solution in a bea-
ker. If possible, use plastic
beakers to minimize any
EMC interference.

CALIBRATION

CALIBRATION

CALIBRATION

CALIBRATION

CALIBRATION

• Immerse the EC or TDS probe in the

solution, making sure that metal pins
are completely submerged.

Note:

 the probe should be submerged ap-

proximately 4 cm (1½”) in the
solution. Moreover, 

i

n order not to

affect the accuracy of measurements,

it is important that probe body does not touch nor
stand close to the side walls of the beaker. The tip
can lay on the bottom of the beaker.

• Wait for a couple of minutes for thermal equilibrium to

be reached.

• Tap the probe gently on the bottom, then shake it

while rotating to make sure no air bubbles have

remained trapped.

• Adjust the calibration trimmer

with the supplied screwdriver

until the display shows
"1.41 mS"  (HI 983302)  or
“1500 ppm” (HI 983301).

• The calibration is now complete and the instrument is

ready for use.

The instrument should be recalibrated at least once a

month and after performing probe cleaning procedure.

ACCESSORIES

ACCESSORIES

ACCESSORIES

ACCESSORIES

ACCESSORIES

HI 7632 *

EC probe with Automatic Temperature Compen-
sation and 2 m (6.6’) cable

HI 7634 *

TDS probe with Automatic Temperature Com-
pensation and 2 m (6.6’) cable

HI 70031P

1413 µS/cm (EC) calibration solution, 20 mL
sachet (25 pcs)

HI 70442P

1500 ppm (TDS) calibration solution, 20 mL

sachet (25 pcs)

HI 7052L

1500 ppm (TDS) calibration sol., 460 mL bottle

HI 7031L

1413 µS/cm(EC) calibration sol., 460 mL bottle

HI 7061L

Electrode cleaning solution, 460 mL bottle

HI 710005

12 VDC power adapter, US plug

HI 710006

12 VDC power adapter, European plug

HI 710012

12 VDC power adapter, Australian plug

HI 710013

12 VDC power adapter, Southern Africa plug

HI 710014

12 VDC power adapter, UK plug

* To be replaced only by authorized service technicians

CE DECLARATION

CE DECLARATION

CE DECLARATION

CE DECLARATION

CE DECLARATION

OF CONFORMITY

OF CONFORMITY

OF CONFORMITY

OF CONFORMITY

OF CONFORMITY

Recommendations for Users

Before using these products, make sure that they are entirely suitable for the environ-

ment in which they are used.
Operation of these instruments in residential areas could cause unacceptable interfer-

ences to radio and TV equipment.
The metal band at the end of the probe is sensitive to electrostatic discharges. Avoid touching

this metal band at all times.
During operation, ESD wrist straps should be worn to avoid possible damage to the probe by

electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s
EMC performance.
To avoid electrical shock, do not use these instruments when voltages at the measurement

surface exceed 24 VAC or 60 VDC.
Use plastic beakers to minimize any EMC interferences.
To avoid damages or burns, do not perform any measurement in microwave ovens.

ISTRECGRR1  09/98