background image

9

GUIDE TO DISPLAY CODES

After Timer is pressed, a blinking hourglass icon appears and the
display shows a 2 minutes coundown. Also the Zero tag might
blink if no zero measurement has been made before. At the end an
acoustic signal alerts the user that the countdown has finished.

Indicates that the instrument is in a ready state and waiting for the
next command (Timer or Zero).

These prompts indicate the type of power supply: “Line” (if the
external power supply is used) or the battery level.

This prompt appears for a few seconds each time the instrument is
turned ON.

The instrument is zeroed and a measurement can be made.

Indicates that the meter is making a measurement.

Batteries voltage is getting low and the batteries need to be
replaced.

Indicates that the meter is performing a zero measurement. The
light intensity is automatically re-adjusted (auto-calibration features)
if necessary.

12

• Whenever the cuvet is placed into the measurement

cell, it must be dry outside, and completely free of
fingerprints, oil or dirt. Wipe it thoroughly with

HI 731318

 (tissue for wiping cuvets, see chapter

ACCESSORIES) or a lint-free cloth prior to insertion.

• Do not let the reacted sample stand too long after

reaction, or accuracy will be lost.

• After the reading it is important to discard immediately

the sample, otherwise the glass might become
permanently stained.

• All the reaction times reported in this manual are

referred to 20°C (68°F). As a general rule of thumb,
they should be doubled at 10°C (50°F) and halved at
30°C (86°F).

Summary of Contents for HI 83741

Page 1: ...Instruction Manual HI 83741 IRON ISM for wine analysis w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives MAN83741R3 04 06 w w w h a n n a i n s t c o m...

Page 2: ...OR AN ACCURATE MEASUREMENT 11 MEASUREMENT PROCEDURE 13 BATTERIES REPLACEMENT 16 ACCESSORIES 16 CE DECLARATION OF CONFORMITY 17 WARRANTY 17 HANNA LITERATURE 18 USER NOTES 19 Dear Customer Thank you for...

Page 3: ...AA batteries Tissue for wiping cuvets Instruction manual Instrument Quality Certificate Rigid carrying case Note save all packing material until you are sure that the instrument works correctly Any de...

Page 4: ...ontent and the type of wine white casse formation white casse formation white casse formation white casse formation white casse formation white casse inhibition white casse inhibition white casse inhi...

Page 5: ...ion suggests good accuracy precise results can be inaccurate The figure explains these definitions In a laboratory using a standard solution of 4 0 mg L iron and a representative lot of reagent an ope...

Page 6: ...n sample and reagents Given that the absorption of a compound strictly depends on the wavelength of the incident light beam a narrow spectral bandwidth should be selected as well as a proper central w...

Page 7: ...y that the surface of the cuvet is clean and not scratched This to avoid measurement interference due to unwanted reflection and absorption of light It is recommended not to touch the cuvet walls with...

Page 8: ...key to perform measurement 8 Power Socket 12V to 20V DC 2 5 Watt FUNCTIONAL DESCRIPTION FRONT REAR INSTRUMENT DESCRIPTION 13 Turn the instrument on by pressing ON OFF MEASUREMENT PROCEDURE When the LC...

Page 9: ...a measurement Batteries voltage is getting low and the batteries need to be replaced Indicates that the meter is performing a zero measurement The light intensity is automatically re adjusted auto ca...

Page 10: ...Please dilute the sample five times see General tips for an accurate measurement page 12 11 The instructions listed below should be carefully followed during testing to ensure best accuracy For a cor...

Reviews: