A microprocessor controlled special tungsten lamp emits radiation which is first optically conditioned
and beamed to the sample contained in the cuvet. The optical path is fixed by the diameter of the
cuvet. Then the light is spectrally filtered to a narrow spectral bandwidth, to obtain a light beam of
intensity 

I

o

 or 

I.

The photoelectric cell collects the radiation 

I

 that is not absorbed by the sample and converts it

into an electric current, producing a potential in the mV range.
The microprocessor uses this potential to convert the incoming value into the desired measuring
unit and to display it on the LCD.
The measurement process is carried out in two phases: first the meter is zeroed and then the actual
measurement is performed.
The cuvet has a very important role because it is an optical element and thus requires particular
attention. It is important that both the measurement and the calibration (zeroing) cuvets are
optically identical to provide the same measurement conditions. Whenever possible use the same
cuvet for both. It is necessary that the surface of the cuvet is clean and not scratched. This to avoid
measurement interference due to unwanted reflection and absorption of light. It is recommended
not to touch the cuvet walls with hands.
Furthermore, in order to maintain the same conditions during the zeroing and the measuring
phases, it is necessary to close the cuvet to prevent any contamination.

US Environmental Protection Agency
degree Celsius
degree Fahrenheit
milligrams per liter. mg/L is equivalent to ppm (part per million)
milliliter
Liquid Crystal Display

EPA

:

°C

:

°F

:

mg/L

:

mL

:

LCD

:

ABBREVIATIONS

14

#2

• Place the cuvet into the holder and close

the lid.

• Press TIMER and the instrument will show the

countdown or, alternatively, wait for 2 minutes.
The instrument gives an acoustic signal to
alert the user that the countdown is finished.

• Press ZERO and “----” will blink on the display.

• Remove the cuvet from the instrument and

open the cap.

• Add the content of one powder packet of

HI 83741B-0 reagent B. Replace the cap and
shake gently for 1 minute to dissolve the
reagent.

• After a few seconds the display will show

“-0.0-”. The meter is now zeroed and ready
for measurement
Note: If the “L Lo” (Low Light) message
appears, the sample must be diluted. See
“General tips for an accurate measurement”
(page 11).

Summary of Contents for HI 83741

Page 1: ...Instruction Manual HI 83741 IRON ISM for wine analysis w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives MAN83741R3 04 06 w w w h a n n a i n s t c o m...

Page 2: ...OR AN ACCURATE MEASUREMENT 11 MEASUREMENT PROCEDURE 13 BATTERIES REPLACEMENT 16 ACCESSORIES 16 CE DECLARATION OF CONFORMITY 17 WARRANTY 17 HANNA LITERATURE 18 USER NOTES 19 Dear Customer Thank you for...

Page 3: ...AA batteries Tissue for wiping cuvets Instruction manual Instrument Quality Certificate Rigid carrying case Note save all packing material until you are sure that the instrument works correctly Any de...

Page 4: ...ontent and the type of wine white casse formation white casse formation white casse formation white casse formation white casse formation white casse inhibition white casse inhibition white casse inhi...

Page 5: ...ion suggests good accuracy precise results can be inaccurate The figure explains these definitions In a laboratory using a standard solution of 4 0 mg L iron and a representative lot of reagent an ope...

Page 6: ...n sample and reagents Given that the absorption of a compound strictly depends on the wavelength of the incident light beam a narrow spectral bandwidth should be selected as well as a proper central w...

Page 7: ...y that the surface of the cuvet is clean and not scratched This to avoid measurement interference due to unwanted reflection and absorption of light It is recommended not to touch the cuvet walls with...

Page 8: ...key to perform measurement 8 Power Socket 12V to 20V DC 2 5 Watt FUNCTIONAL DESCRIPTION FRONT REAR INSTRUMENT DESCRIPTION 13 Turn the instrument on by pressing ON OFF MEASUREMENT PROCEDURE When the LC...

Page 9: ...a measurement Batteries voltage is getting low and the batteries need to be replaced Indicates that the meter is performing a zero measurement The light intensity is automatically re adjusted auto ca...

Page 10: ...Please dilute the sample five times see General tips for an accurate measurement page 12 11 The instructions listed below should be carefully followed during testing to ensure best accuracy For a cor...

Reviews: