40
MAINTENANCE
13.2. PUMP TUBING REPLACEMENT
Note:
While replacing tubing, wear protective gloves and eye protection at all times.
1. Power off the controller.
2B
3C
2A
2. Remove the plastic screw securing the transparent cover (2A) and the cover (2B).
3. Disconnect the tubing from the pump (3C).
4. Starting from the left side of the pump, grab the peristaltic pump tubing (4A) and rotate the
pump rotor manually to the right (4B), until the tubing is removed (4C).
5. Grease the new peristaltic pump tubing with silicone oil supplied in the peristaltic pump tubing
kit (
BL100-300
).
4B
4C
4A