22
SETUP
7. SETUP
• Short press MENU key to move to next item in menu.
• Long press MENU key to exit.
• Press arrow keys to change the values.
• Short press MENU key to automatically save modified values.
Table below presents an overview of the menu with ranges and factory set defaults.
Parameter
(Scrolled Message)
Range / Option
Default Settings
Control
Auto or oFF
Auto
Control type
on/oFF or ProP (Proportional)
on/oFF
Control mode
Hi (High) or Lo (Low)
Lo (Low)
Measurement mode
EC or tdS
EC
Set point
0.10 to 10.00 mS/cm
45 to 9900 ppm (depends on TDS factor)
2.50 mS/cm
1250 ppm
Hysteresis (on/oFF only)
0.05 to 0.50 mS/cm
23 to 495 ppm (depends on TDS factor)
0.10 mS/cm
50 ppm
Band (Proportional only)
0.05 to 1.00 mS/cm
23 to 990 ppm (depends on TDS factor)
0.10 mS/cm
50 ppm
Start up delay (seconds)
0 to 600
60
Overtime alarm (minutes) oFF, 1 to 180
30
Flow rate (Liter/hour)
Flow rate (Gallon/hour)
0.5 to 3.5
0.13 to 0.92
1.0
0.26
Level alarm
diS (disabled) or En (enabled)
diS (disabled)
High alarm
diS (disabled) or En (enabled)
diS (disabled)
High alarm value
0.10 to 10.00 mS/cm*
10 to 9900 ppm (depends on TDS factor)*
9.50 mS/cm
4750 ppm
Low alarm
diS (disabled) or En (enabled)
En (enabled)
Low alarm value
0.00 to 9.90 mS/cm*
0 to 9890 ppm (depends on TDS factor)*
0.50 mS/cm
250 ppm
TDS factor (TDS mode only) 0.45 to 0.99
0.50
Temperature coefficient
0 to 2.4% / °C
1.9% / °C
Temperature unit
°C or °F
°C
Flow rate unit
L.H (Liter/hour) or GAL.H (Gallon/hour)
L.H (Liter/hour)