32
EVENT MANAGEMENT
9.2. WARNINGS
Two types of warnings can be independently enabled or disabled in SETUP. If any of the warnings is
active, the LCD backlight turns yellow.
Warnings
Description
Screenshot for
warning solution
Terminate condition
Startup delay Start up delay is active
Press the up / down
arrows at same time to
restart pump.
Control Off
Control is switched Off
Set Control setting to
Auto.
9.3. SUMMARY OF GENERAL BEHAVIOR
Behavior
Control
Backlight
Alarm Output Relay
Control - auto
auto
green
off
Warning active
off
yellow
off
Error active
off
red
on
Alarm active
off
red (blinking)