SETUP
Option:
user selectable
Press the arrow keys to set the value.
“HYSTERESIS” or “PROPORTIONAL BAND” message is scrolled on the bottom of the LCD screen.
Note:
To enter Hysteresis / Band screen, pump control mode must be set as Auto.
Startup Delay (Automatic Control Only)
Option:
user selectable (0 to 600s)
Startup delay represents the delay to start dosing at power-on.
Press the arrow keys to change the time values.
“STARTUP DELAY SEC” message is scrolled on the bottom of the LCD screen.
Note:
To enter Startup Delay screen, pump control mode must be set as Auto.
Overtime Alarm
Option:
user selectable (off, 1 to 180 minutes)
Press the arrow keys to change the time values.
To disable the alarm, select Off.
“OVERTIME ALARM MIN” message is scrolled on the bottom of the LCD screen.
Note:
To enter Overtime alarm screen, pump control mode must be set as Auto.