1. Safety Precautions
1.1 Safety requirements
Centrifuge involves elements of risk all the time as it utilizes a high speed body of rotation. Follow precautions and all the safety
Follow precautions and all the safety requirements described on this user manual to prevent any damage and failure of equipment
and loss of lives.
1. Centrifuge should be placed on flat place to maintain level. Big vibration is caused by operation of equipment in the condition
of spindle inclined.
2. Check voltage to be used before centrifuge is connected to power supply.
Wrong voltage can damage equipment.
3. Use rotors provided by Hanil Science Industrial Co Ltd and use parts and accessories recommended.
Hanil Science Industrial Co Ltd is not responsible for damages of equipment and accidents caused by using parts and accessories
not recommended.
4. Understand property of matter fully and take needful safety actions in case you perform centrifugation of dangerous materials
such as pathogenic, toxic and radioactive substances.
5. Remove contaminants completely and take needful actions such as ventilation or isolation of centrifuge in case centrifuge is
contaminated with dangerous materials such as pathogenic, toxic and radioactive substances.
6. Do not perform centrifugation of materials capable of producing volatile and explosible steam.
7. Do not operate equipment more than maximum RPM allowed by rotor installed.
8. Do not decelerate or stop rotor rotating with hands.
9. Rotor should be used after balancing of symmetry between weight of samples to be used.
10. Rotor should be fixed into shaft and its lid should be tightened completely.
Otherwise, the lid can be taken off during rotation.
11. A trained and qualified technician should perform repairing directions not mentioning on this user manual.
12. User should remove contaminants completely in advance in case user requests repairs to a qualified technician.
13. Rotor and chamber should be kept dry all the time before use of centrifuge to make long period use of centrifuge and maintain
the best operating conditions.
14. Pay attention that aluminum rotor has a chemical reaction and starts to be corrosive if aluminum rotor contacts laboratory
cleaning liquid, cesium, silver, and (especially chloride ion) of strong acid, strongly alkaline or strong alkali.
15. According to IEC61010-2-020, 30cm of a safe distance around centrifuge should be assured during operation.
Nothing should be put on this safe distance in case it is destroyed.