5. Troubleshooting
5.2 Emergency lid open system
In case a lid closed due to power failure, activate emergency lid open system manually.
1.Switch off main power. Wait till a rotor stops through RPM measuring window.
2.As the below picture, Insert a flathead screwdriver into emergency lib open system on the left side of
the equipment.
3. Turn the flathead screwdriver clockwise to open the lid.
www.ihanil.com