General information regarding the CE marking

HAMEG instruments fulfi ll the regulations of the EMC directive. The 

conformity test made by HAMEG is based on the actual generic- and 

product standards. In cases where different limit values are applicable, 

HAMEG applies the severer standard. For emission the limits for 

residential, commercial and light industry are applied. Regarding the 

immunity (susceptibility) the limits for industrial environment have 

been used.

The measuring- and data lines of the instrument have much infl uence 

on emission and immunity and therefore on meeting the acceptance 

limits. For different applications the lines and/or cables used may 

be different. For measurement operation the following hints and 

conditions regarding emission and immunity should be observed:  

1. Data cables

For the connection between instrument interfaces and external devices, 

(computer, printer etc.) suffi ciently screened cables must be used. 

Without a special instruction in the manual for a reduced cable length, 

the maximum cable length of a dataline must be less than 3 meters and 

not be used outside buildings. If an interface has several connectors 

only one connector must have a connection to a cable.

Basically interconnections must have a double screening. For IEEE-bus 

purposes the double screened cables HZ73 and HZ72L from HAMEG 

are suitable.

2. Signal cables

Basically test leads for signal interconnection between test point and 

instrument should be as short as possible. Without instruction in the 

manual for a shorter length, signal lines must be less than 3 meters 

and not be used outside buildings.

Signal lines must screened (coaxial cable - RG58/U). A proper ground 

connection is required. In combination with signal generators double 

screened cables (RG223/U, RG214/U) must be used.

3. Infl uence on measuring instruments

Under the presence of strong high frequency electric or magnetic 

fi elds, even with careful setup of the measuring equipment, infl uence 

of  such signals is unavoidable.

Die HAMEG Instruments GmbH bescheinigt die Konformität für das Produkt
The HAMEG Instruments GmbH herewith declares conformity of the product 
HAMEG Instruments GmbH déclare la conformite du produit

Bezeichnung / Product name / Designation: 
  

 

Oszilloskop

  

 

Oscilloscope

  

 

Oscilloscope

Typ / Type / Type: 

 

HM1008-2

mit / with / avec:     

Optionen / Options / Options:   

mit den folgenden Bestimmungen / with applicable regulations / avec les 
directives suivantes

 

EMV Richtlinie 89/336/EWG ergänzt durch 91/263/EWG, 92/31/EWG

 

EMC Directive 89/336/EEC amended by 91/263/EWG, 92/31/EEC

 

Directive EMC 89/336/CEE amendée par 91/263/EWG, 92/31/CEE

Niederspannungsrichtlinie 73/23/EWG ergänzt durch 93/68/EWG
Low-Voltage Equipment Directive 73/23/EEC amended by 93/68/EEC
Directive des equipements basse tension 73/23/CEE amendée par 93/68/CEE

Angewendete harmonisierte Normen / Harmonized standards applied / Normes 
harmonisées utilisées:

Sicherheit / Safety / Sécurité: EN 61010-1:2001 (IEC 61010-1:2001)
Überspannungskategorie / Overvoltage category / Catégorie de surtension: II
Verschmutzungsgrad / Degree of pollution / Degré de pollution: 2

Elektromagnetische Verträglichkeit / Electromagnetic compatibility / 
Compatibilité électromagnétique

EN 61326-1/A1 Störaussendung / Radiation / Emission: 
 

Tabelle / table / tableau 4; Klasse / Class / Classe B.

Störfestigkeit / Immunity / Imunitée: Tabelle / table / tableau A1. 

EN 61000-3-2/A14 Oberschwingungsströme / Harmonic current emissions / 
Émissions de courant harmonique: 
 

Klasse / Class / Classe D.

EN 61000-3-3 Spannungsschwankungen u. Flicker / Voltage fl uctuations and fl icker / 
Fluctuations de tension et du fl icker.

Datum /Date /Date                                          
01. 06. 2007
 

 

 

Unterschrift / Signature / Signatur 

  

 

 

Holger 

Asmussen

  

 

 

Manager

This will not cause damage or put the instrument out of operation. Small 

deviations of the measuring value (reading) exceeding the instruments 

specifi cations may result from such conditions in individual cases.

4. RF immunity of oscilloscopes.

4.1 Electromagnetic RF fi eld

The infl uence of electric and magnetic RF fi elds may become visible 

(e.g. RF superimposed), if the fi eld intensity is high. In most cases 

the coupling into the oscilloscope takes place via the device under 

test, mains/line supply, test leads, control cables and/or radiation. 

The device under test as well as the oscilloscope may be effected by 

such fi elds.

Although the interior of the oscilloscope is screened by the cabinet, 

direct radiation can occur via the CRT gap. As the bandwidth of 

each amplifi er stage is higher than the total –3dB bandwidth of the 

oscilloscope, the infl uence of RF fi elds of even higher frequencies 

may be noticeable. 

4.2 Electrical fast transients / electrostatic discharge

Electrical fast transient signals (burst) may be coupled into the 

oscilloscope directly via the mains/line supply, or indirectly via test 

leads and/or control cables. Due to the high trigger and input sensitivity 

of the oscilloscopes, such normally high signals may effect the trigger 

unit and/or may become visible on the CRT, which is unavoidable. 

These effects can also be caused by direct or indirect electrostatic 

discharge.  

   

HAMEG Instruments GmbH

 Hersteller 

HAMEG 

Instruments 

GmbH  KONFORMITÄTSERKLÄRUNG 

 

 

Manufacturer 

Industriestraße 6 

DECLARATION OF CONFORMITY 

 

Fabricant 

D-63533 Mainhausen 

DECLARATION DE CONFORMITE

Summary of Contents for HM1008-2

Page 1: ...100 MHz CombiScope HM1008 2 Manual English 99 Washington Street Melrose MA 02176 Fax 781 665 0780 TestEquipmentDepot com...

Page 2: ...e 89 336 EEC amended by 91 263 EWG 92 31 EEC Directive EMC 89 336 CEE amend e par 91 263 EWG 92 31 CEE Niederspannungsrichtlinie 73 23 EWG erg nzt durch 93 68 EWG Low Voltage Equipment Directive 73 23...

Page 3: ...ion 14 Phase measurements with Lissajous gures 14 Measurement of phase differences in dual channel Yt mode 14 Measurement of amplitude modulation 15 Triggering and time base 15 Automatic peak triggeri...

Page 4: ...Flash A D Converters Pre Post Trigger 100 to 400 Time Base 50 s cm 5 ns cm Acquisition modes Single Event Refresh Average Envelope Roll Peak Detect USB RS 232 optional IEEE 488 Ethernet USB Signal dis...

Page 5: ...Hz 3 dB X Y phase shift 3 220 kHz Digital mode Time base range 1 2 5 sequence Refresh Mode 20 ms cm 5 ns cm with Peak Detect 20ms cm 2ms cm min Pulse Width 10ns Roll Mode 50 s cm 50 ms cm Accuracy tim...

Page 6: ...as manufactured and tested accordingly It left the factory in perfect safe condition Hence it also corresponds to European Standard EN 61010 1 resp International Standard IEC 1010 1 In order to mainta...

Page 7: ...0 to 40 degrees C During transport or storage the temperature may be 20 to 55 degrees C Please note that after exposure to such temperatures or in case of condensation proper time must be allowed unti...

Page 8: ...d 2 controls 16 VOLTS DIV SCALE VAR knob 34 Channel 1 Y de ection coef cient Y variabel and Y scaling setting 17 VOLTS DIV SCALE VAR knob 34 Channel 2 Y de ection coef cient Y variabel and Y scaling s...

Page 9: ...e rear side and USB Stick ash drive connector 41 COMPONENT TESTER 2 sockets with 4 mm 47 Connectors for test leads of the Component Tester Left socket is galvanically connected with protective earth 4...

Page 10: ...owever when charging and discharging a large capacitor DC coupling is preferable with all signals of varying duty cycle otherwise the display will move up and down depending on the duty cycle Of cours...

Page 11: ...ignal to 5 cm Position the rising portion symmetrically to the graticule centre line using both Y and X positioning controls Notice the intersections of the signal with the 10 and 90 lines and project...

Page 12: ...r other interference is visible especially at high sen sitivity one possible reason may be multiple grounding The scope itself and most other equipment are connected to safety ground so ground loops m...

Page 13: ...r with both a high CM rejection and a high permissible CM range Therefore please observe the following rule Always look at the two signals in the one channel only or the dual modes not in Add mode and...

Page 14: ...ometric functions are periodic limit the cal culation to angles 90 degrees This is where this function is most useful Do not use too high frequencies because as explained above the two amplifiers are...

Page 15: ...to be found in the shaded TRIGGER area they are described in Controls and Readout In YT mode the signal will de ect the trace vertically while the ti mebase will de ect it horizontally the speed can...

Page 16: ...automatic and normal modes Trigger coupling Menu FILTER Consult chapters MODE 20 AUTO LEVEL A B 19 FILTER 21 and SOURCE 22 in Controls and Readout In AUTOSET DC coupling will be used unless ac couplin...

Page 17: ...rds will indicate the trigger time posi tion if this lies within the screen area The Trigger symbol is not indicated This trigger mode is to be used with greatest care and should be an exception rathe...

Page 18: ...a reconstruction of the signal acquired In analog mode the display will always start on the left Let us assume one period of a signal is displayed at a convenient time base setting Increasing the swe...

Page 19: ...COMPONENT TESTER 41 The scope has a built in component tester The test object is con nected with 4 mm banana plugs In this mode the Y ampli ers and the time base are turned off If the components are...

Page 20: ...part is good With enhancement type MOSFETs an open circuit will be seen in all directions as the threshold voltage G S is not available With power MOSFETs the antiparallel diode S D can be seen Tests...

Page 21: ...the sampling frequency is only twice the signal frequency there will only be two points displayed on the screen any number of signal shapes may be drawn which t through these two points The Nyquist th...

Page 22: ...over the screen from left to right in usual Yt mode Roll readout rol 3 Single sweep triggered menu Trigger MODE signal capture in usual Yt mode Single readout sgl 4 Untriggered continuous signal captu...

Page 23: ...en exactly This is dependent on the signal shape as well as on its amplitude displayed While it is fairly easy to recognize a square wave it requires at least 10 samples per period to distinguish a si...

Page 24: ...le a communication between PC and oscilloscope an USB driver must be installed on the PC which is requested by Windows if for the rst time a connection is made between oscilloscope and PC via USB A di...

Page 25: ...e settings Also arrows may be shown which point to available submenus In some modes various pushbuttons or INTENS operations are meaningless and will hence not cause a menu display STOP Please note If...

Page 26: ...the signal expanded by Zoom RO Int Readout intensity Focus Focus for signal and readout Readout On Off The readout can be switched off to eliminate interference of the readout with the signal s that...

Page 27: ...determined by the distance of the short auxiliary cursor line to the long reference line lower respectively left if appropriate with a negative sign 4 2 1 3 degree measurement of degrees The distance...

Page 28: ...7 MATH pushbutton digital mode only not in FFT mode The MATH pushbutton calls the Mathematics menu and a for mula editor The mathematic menu enables the mathematical processing of current channel 1 an...

Page 29: ...opped pushbutton lit or started pushbutton not lit by the RUN STOP pushbutton 6 A trigger will start a new acquisition which will overwrite the display of the former The display will remain on screen...

Page 30: ...ween the samples and short signal deviations such as glitches may occur without being sampled The maximum gap size is 0 25 s between the samples at a time base setting of 50 s cm The advantage of Peak...

Page 31: ...ssocia Set STOP Please note If both reference displays are on and if both memory locations are identical e g RE1 RE1 the signal will be displayed twice on the same spot 9 2 2 4 Associated settings Pre...

Page 32: ...f a menu was opened it will be turned off by AUTOSET During the display of HELP texts AUTOSET is not available AUTOSET can and should be used during FFT ope ration to avoid misadjustment 12 HELP Press...

Page 33: ...shbutton 15 and the menu item acti vated 14 1 Y position 14 1 1 Y Position of channel 2 analog and digital mode POSITION 2 is for Y Position CH2 control if Yt mode time base mode is present and the CH...

Page 34: ...low too many signal periods to avoid so called aliasing The signal display height should be between 5 mm 0 5 div and 8 cm 8 div Signal display heights 8 cm cause the danger that the dynamic range is e...

Page 35: ...In principle the following conditions must be met a Trigger conditions must be present if frequency or period time measurement is performed For signals under 20 Hz normal triggering is required Please...

Page 36: ...ep triggering can be selected Choo sing Slope will allow trigger on any signal shape For video signals select Video and press the FILTER pushbutton 21 in order to nd a choice of special trigger modes...

Page 37: ...ngs are available 21 2 1 Frame Line Depending on the setting chosen triggering will be on frame or line sync pulses The selection will also affect other menu items Readout Tr Source TV 21 2 1 1 Frame...

Page 38: ...This pushbutton allows you to change the function of the HO RIZONTAL knob 27 26 1 Analog mode The pushbutton signals the actually selected function in accor dance with the front panel lettering dark...

Page 39: ...he exception of 50 ns cm TB B can not be set to the same speed as TB A or slower Further information is available in the section Time base B 2nd time base Delay Triggering Analog mode in the chapter T...

Page 40: ...th other parameters in FFT mode From the sampling frequency rate setting the theoretically highest displayable frequency fmax results In accordance with the Nyquist Shannon sampling theorem the highes...

Page 41: ...resentation by time base A there will be no triggering of time base B any more and thus no time base B display The same holds in time base B only mode 30 1 5 B trigger Edge Except for the negative edg...

Page 42: ...ls with low frequency content If the DC content of the signal or the duty cycle of pulses changes the capacitor will charge or discharge and this will cause a momentary Y shift of the display 31 2 Gro...

Page 43: ...ng is necessary and no information pertaining to it is shown FFT DUAL mode will not be offered in combination with FFT 32 4 ADD In ADD mode the signals of both channels are algebraically added and dis...

Page 44: ...tion of it e g to the screen centre without disconnecting the signal After switching back to the signal its amplitude can now be determined with respect to the formerly set zero reference The 0 V Grou...

Page 45: ...uency is the highest frequency a real time digital oscilloscope can measure without aliasing This frequency is half the sampling frequency sam pling rate In case of higher frequencies the number of sa...

Page 46: ...bmenus Save and Recall that can be opened by pressing the assigned function keys When the menu Utilities is not displayed it opens automati cally when a USB ash drive is connected with the USB Stick c...

Page 47: ...on can be found under section Component Tester 42 USB Stick USB ash drive connector After a USB ash drive has been connected the LED in the USB ash drive brie y blinks and the menu Utilities opens Fur...

Reviews: