
Option
Description
Verify after Cal.
Sets the instrument to start a verification immediately after the instrument is calibrated.
When set to on, the verification standard is measured immediately after a calibration is
done. Default: ON.
Acceptance
Range
Sets the maximum difference permitted between the recorded value of the verification
standard and the measured value of the verification standard during verification.
Unit
—
Sets the acceptance range for verification to a percentage (1 to 99%) or an NTU value
(0.015 to 100.00 NTU). Options: % or NTU (or mNTU).
Verification
Reminder
Sets the time interval between calibration verifications. The display will show a reminder
when a verification is due. Options: OFF(default), 1 day, 7 days, 30 days or 90 days.
When a verification is done, the verification time is set to zero.
Do a calibration verification
Do a calibration verification to make sure that the instrument is still calibrated as necessary.
1.
Log in to the instrument as follows:
• Put an operator RFID tag in front of the RFID module or
• Push
Login
. Select the applicable operator ID, then push
Select
.
2.
Push
Verification
.
3.
Clean the verification standard vial with a no-lint cloth to remove contamination.
Keep contamination off the glass. Refer to
Prevent vial contamination
on page 18.
4.
Dry the external surfaces of the vial with a no-lint cloth. Make sure to dry the bottom of the vial.
5.
Carefully invert the vial three times. Refer to
Figure 6
on page 27.
6.
Put the vial in the vial compartment.
7.
Push
Start
.
8.
Confirm the value shown or enter the exact turbidity value of the standard. Push
Read
.
9.
When the measurement is complete, remove the vial.
The verification is recorded to the verification log.
10.
To show the verification log, push
Options
>
Verification Log
. Refer to
Show the recorded data
on page 24 for more options.
11.
To send the verification data to external devices that are connected to the instrument, push
Options
>
Send Data
. Refer to
Show the recorded data
on page 24 for more options.
Maintenance
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
28
English
Summary of Contents for TU5200
Page 1: ...DOC342 53 80489 TU5200 03 2017 Edition 3 User Manual...
Page 2: ......
Page 38: ...36 English...
Page 39: ......