Hach TU5200 User Manual Download Page 30

Option

Description

Verify after Cal.

Sets the instrument to start a verification immediately after the instrument is calibrated.
When set to on, the verification standard is measured immediately after a calibration is
done. Default: ON.

Acceptance
Range

Sets the maximum difference permitted between the recorded value of the verification
standard and the measured value of the verification standard during verification. 

Unit

Sets the acceptance range for verification to a percentage (1 to 99%) or an NTU value
(0.015 to 100.00 NTU). Options: % or NTU (or mNTU).

Verification
Reminder

Sets the time interval between calibration verifications. The display will show a reminder
when a verification is due. Options: OFF(default), 1 day, 7 days, 30 days or 90 days.
When a verification is done, the verification time is set to zero.

Do a calibration verification

Do a calibration verification to make sure that the instrument is still calibrated as necessary.

1.

Log in to the instrument as follows:

• Put an operator RFID tag in front of the RFID module or
• Push 

Login

. Select the applicable operator ID, then push 

Select

.

2.

Push 

Verification

.

3.

Clean the verification standard vial with a no-lint cloth to remove contamination.
Keep contamination off the glass. Refer to 

Prevent vial contamination

 on page 18.

4.

Dry the external surfaces of the vial with a no-lint cloth. Make sure to dry the bottom of the vial.

5.

Carefully invert the vial three times. Refer to 

Figure 6

 on page 27.

6.

Put the vial in the vial compartment.

7.

Push 

Start

.

8.

Confirm the value shown or enter the exact turbidity value of the standard. Push 

Read

.

9.

When the measurement is complete, remove the vial.
The verification is recorded to the verification log.

10.

To show the verification log, push 

Options

>

Verification Log

. Refer to 

Show the recorded data

on page 24 for more options.

11.

To send the verification data to external devices that are connected to the instrument, push

Options

>

Send Data

. Refer to 

Show the recorded data

 on page 24 for more options.

Maintenance

C A U T I O N

 

Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.

C A U T I O N

 

Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.

28

   

English

Summary of Contents for TU5200

Page 1: ...DOC342 53 80489 TU5200 03 2017 Edition 3 User Manual...

Page 2: ......

Page 3: ...perator IDs 13 Configure an operator RFID tag optional 14 Add sample IDs 14 Import sample IDs optional 14 Configure the measurement settings 15 Set the acceptance range 15 Set up a LAN connection 16 C...

Page 4: ...gs 27 Do a calibration verification 28 Maintenance 28 Clean spills 29 Clean the instrument 29 Clean a sample vial 29 Clean the vial compartment 30 Troubleshooting 30 Remove air bubbles from the sample...

Page 5: ...EBC Accuracy 2 of reading plus 0 01 NTU from 0 to 40 FNU 10 of reading from 40 to 1000 FNU based on Formazin primary standard at 25 C 77 F Linearity Better than 1 for 0 to 40 NTU on Formazin at 25 C...

Page 6: ...ims such damages to the full extent permitted under applicable law The user is solely responsible to identify critical application risks and install appropriate mechanisms to protect processes during...

Page 7: ...electrocution exists This symbol indicates the need for protective eye wear This symbol indicates a laser device is used in the equipment This symbol identifies a risk of chemical harm and indicates...

Page 8: ...se of doubt contact the manufacturer Safety information for RFID modules W A R N I N G Multiple hazards Do not disassemble the instrument for maintenance If the internal components must be cleaned or...

Page 9: ...ncy energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to...

Page 10: ...rt for LAN connection 4 Power button 9 RFID module indicator optional 5 Power supply connection 10 USB port type A Product components Make sure that all components have been received Refer to Figure 2...

Page 11: ...lation to break down which can result in an electric shock hazard The manufacturer recommends that users with concerns contact technical support Installation guidelines Install the instrument On a lev...

Page 12: ...o not use writing tips of pens or pencils or other sharp objects to make selections on the screen or damage to the screen will occur Refer to Figure 3 for an overview of the home screen Figure 3 Displ...

Page 13: ...ading log calibration log verification log and compare log Refer to Show the recorded data on page 24 Setup Configures the instrument settings Refer to Configure the instrument settings on page 12 Dia...

Page 14: ...mmm yyyy default yyyy mm dd dd mm yyyy or mm dd yyyy Time Format Sets the time format Options 12 or 24 hours default Security Enables or disables password protection for the settings and tasks in the...

Page 15: ...nguage and then push OK Add operator IDs Add a unique operator ID for each person who will measure samples 30 maximum Select an icon operator password and security level for each operator ID 1 Push Lo...

Page 16: ...e barcode with a barcode handset scanner that is connected to the instrument The barcode is added to the sample ID 5 Push OK 6 Select an option Option Description Add Date Time Adds the data and time...

Page 17: ...s the measurement units that show on the display and that are recorded to the data log Options NTU FNU TE F FTU EBC mNTU or mFNU Default FNU Data Log Setup Sets the data log settings Auto Store Measur...

Page 18: ...Select Network then push Setup 5 Select On 6 Push TU5200 IP 7 To get an IP address for the instrument from a DHCP server do the steps that follow a Select Automatic b Push OK two times The IP address...

Page 19: ...network printers that are SNMP enabled and in the same sub network are found by the instrument 1 Push two times then push Setup 2 Push Network Peripherals 3 Select Printer then push Setup 4 Push Netwo...

Page 20: ...th the sample When collecting a sample from a water tap in a distribution system or treatment plant turn the water on for at least five minutes then collect the sample Do not adjust the flow because t...

Page 21: ...l and national regulations N O T I C E Always put a cap on the sample vial to prevent spills in the vial compartment Refer to the illustrated steps that follow to prepare a sample vial for measurement...

Page 22: ...operator RFID tag in front of the RFID module or Push Login Select the applicable operator ID then push Select 2 Select the sample ID as follows Put the sample RFID sticker on the sample bottle in fro...

Page 23: ...cess and lab instrument are calibrated with the same number of calibration points and with the same standards Make sure that the calibrations are not expired Collect a grab sample Collect a 100 mL sam...

Page 24: ...eading from the process instrument shows on the display 11 Prepare a grab sample vial Refer to Prepare a sample vial on page 19 12 Measure the turbidity of the grab sample with the laboratory instrume...

Page 25: ...e laboratory instrument 7 Move the grab sample bottle to the laboratory instrument 8 On the TU5200 push Options Reading Setup 9 Push Bubble Reject then set bubble reject to on 10 If the grab sample is...

Page 26: ...Calibration log Shows the calibration history Verification log Shows the verification history Compare log Shows the recorded comparisons of process and laboratory measurements 1 Push Data Log and sel...

Page 27: ...data point Data points Selects the symbol used for the data points Control Limit Sets the minimum value and maximum value of the readings that show on the graph Show the instrument information 1 Push...

Page 28: ...er with kaolin SDVB 0 1000 FNU 3 point calibration 20 FNU 600 FNU and dilution water with spherical slyrene divinylbenzene Custom Calibration 1 to 6 point calibration 0 02 to 1000 FNU with StablCal or...

Page 29: ...measured 12 Push Store to complete the calibration and record the results to the calibration log 13 To show the calibration log push Options Calibration Log Refer to Show the recorded data on page 24...

Page 30: ...ect 2 Push Verification 3 Clean the verification standard vial with a no lint cloth to remove contamination Keep contamination off the glass Refer to Prevent vial contamination on page 18 4 Dry the ex...

Page 31: ...current safety data sheets MSDS SDS for safety protocols Clean the sample vial when there is contamination in the sample vial after the sample vial is rinsed Items to collect Hydrochloric acid concen...

Page 32: ...micro fiber cloth lint free cloth Liquid oil Cloth water and cleaning agent Troubleshooting Message Solution Startup The self check stopped Clean the vial compartment The vial compartment is dirty Set...

Page 33: ...stable Use the correct calibration standards Invert the standard until no bubbles or large particles show The standard value is out of the measurement range Use the correct calibration standards Inve...

Page 34: ...that the file system FAT32 is installed on the USB flash drive Set the power to off wait 20 seconds and then set the power to on again Look for remaining space on the USB flash drive Set the power to...

Page 35: ...k administrator If the DHCP server is not enabled manually enter an IP address for the instrument Setting IP address failed Contact the network administrator If necessary contact technical support Set...

Page 36: ...ollects on the sample vial during measurement let the sample become slightly warmer before measurement Keep the sample at room temperature or partially put the sample vial into a warm water bath for a...

Page 37: ...60 Hz 1 LZV798 Sample vials with caps 2 LZV946 Vial stand 1 LZV952 Accessories Description Quantity Item no Micro fiber cloth vial cleaning 1 LZY945 RFID tags operator 2 pkg LZQ066 RFID stickers samp...

Page 38: ...36 English...

Page 39: ......

Page 40: ...orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 route de...

Reviews: