1;Aeration;13.09.2015 10:03
2;Outlet;13.09.2015 06:30
3;Feed;13.09.2015 18:00
Note: The date of sample collection is optional.
4.
Make a new folder on a USB flash drive. Give the folder the name "SampleID".
5.
Save the spreadsheet file to the SampleID folder as a CSV (comma-separated value) or TXT
(text) file.
6.
Connect the USB flash drive to a USB port on the instrument.
7.
On the instrument, push
Sample ID
>
Options
>
Import Sample ID list
.
The filename of the spreadsheet file(s) in the SampleID folder shows.
8.
Select the applicable spreadsheet file, then push
OK
.
The sample IDs are added to the instrument.
Configure the measurement settings
Select the reading mode, measurement units, data log settings, resolution and more.
1.
At the main reading screen, push
Options
>
Reading Setup
.
2.
Select an option.
Option
Description
Reading
Sets the reading mode to single, continuous or minimum mode. Default: Single.
Single
—
The measurement stops when the reading is stable.
Continuous
—The measurement
continues until the user pushes
Done
.
Minimum Mode
—Set to on when a process and
laboratory measurement are compared and the process measurement is a lower NTU
range. Removes the effect of non-representative particles in the grab sample.
Signal Avg
—The turbidity reading that shows on the display is an average of the values measured
during the time interval selected. Options: For single measurement mode, 5 to 15 seconds.
For continuous measurement mode, 5 to 90 seconds.
Unit
Selects the measurement units that show on the display and that are recorded to the data
log. Options: NTU, FNU, TE/F, FTU, EBC, mNTU or mFNU. Default: FNU).
Data Log
Setup
Sets the data log settings.
Auto Store
—Measurement data is automatically recorded in the
reading log. Default: On. When not selected, push
Options
>
Store
to record the current
measurement to the reading log as necessary.
Send Data Format
—Sets the output format
of measurement data that is sent to external devices (CSV or XML). Default: XML.
Format
—Sets the output format of measurement data that is sent to a printer (Quick Print
or Detailed Print (GLP)).
Comments
—Lets users add comments to log entries.
Auto Send
—Measurement data is automatically sent to all of the devices (e.g., printer, USB flash
drive and FTP server) that are connected to the instrument after each measurement.
Resolution
Selects the number of decimal places that show on the display. Options: 0.001 (default) or
0.0001.
Bubble Reject
Sets the bubble reject to on (default) or off. When set to on, high turbidity readings caused
by bubbles in the sample are not shown or saved to the data log.
Close lid to
start reading
Enables or disables the instrument to start a measurement automatically when the lid is
closed. Default: On. A measurement is only done when there is a sample vial in the
instrument.
Set the acceptance range
Before process and laboratory measurements are compared on the instrument, set the acceptance
range for the compare results. The acceptance range is the maximum difference permitted between
the process and laboratory measurements.
1.
Push
LINK2SC
.
2.
Push
Options
>
Compare Setup
.
3.
Push
Acceptance Range
>
Unit
.
English
15