Hach TU5200 User Manual Download Page 24

• Glass sample bottle with a sample RFID sticker
• Operator RFID tag (optional)

1.

At the process instrument, put the operator RFID tag (if available) near the RFID module.

2.

Put a sample RFID sticker on the sample bottle.

3.

Collect a grab sample. Refer to 

Collect a grab sample

 on page 21.

4.

At the process instrument, put the RFID sticker that is on the sample bottle near the RFID
module.
The instrument gives a sound signal. The status indicator light changes to blue.
The turbidity reading, operator ID (if available), location of the process instrument and the date
and time are recorded on the RFID sticker.

5.

Move the grab sample bottle to the laboratory instrument.

6.

On the TU5200, push 

Options

>

Reading Setup

.

7.

Push 

Bubble Reject

, then set bubble reject to on.

8.

If the grab sample is 1 NTU or less, push 

Reading

>

Minimum Mode

, then select 60 seconds.

Note: In minimum mode, readings are done continuously for 60 seconds when a measurement is done. The
smallest reading within 60 seconds is saved to the data log.

9.

At the laboratory instrument, put the operator RFID tag (if available) near the RFID module to log
in. Refer to 

Figure 1

 on page 8 for the location of the RFID module.

10.

Put the RFID sticker that is on the sample bottle near the RFID module.
The instrument gives a sound signal. The turbidity reading from the process instrument shows on
the display.

11.

Prepare a grab sample vial. Refer to 

Prepare a sample vial

 on page 19.

12.

Measure the turbidity of the grab sample with the laboratory instrument. Refer to the
TU5200 documentation.
If the difference between the process and laboratory measurements is not more than the selected
acceptance range, "Measurement values match." shows on the display. Refer to 

Set the

acceptance range

 on page 15 to select the acceptance range.

If "Measurement values do not match." shows on the display, click the link to show the
troubleshooting steps.

13.

To show the compare log, push 

Options

>

Compare Log

. Refer to the TU5200 documentation for

more options.

14.

To send the verification data to external devices that are connected to the instrument, push

Options

>

Send Data

. Refer to the TU5200 documentation for more options.

Compare measurements with Link2SC

When the process instrument and laboratory instrument do not have the optional RFID module,
compare the process and laboratory measurements with Link2SC.

Items to collect:

• TU5300 sc or TU5400 sc
• TU5200 
• TU5200 sample vials
• SD card

5

 (or a LAN connection at the SC controller

6

 and the laboratory instrument

7

)

• USB adapter for the SD card (if used)

5

Refer to the SC controller documentation for the SD card requirements.

6

Refer to the SC controller documentation to set up a LAN connection at the SC controller.

7

Refer to the TU5200 documentation to set up a LAN connection at the laboratory instrument.

22

   

English

Summary of Contents for TU5200

Page 1: ...DOC342 53 80489 TU5200 03 2017 Edition 3 User Manual...

Page 2: ......

Page 3: ...perator IDs 13 Configure an operator RFID tag optional 14 Add sample IDs 14 Import sample IDs optional 14 Configure the measurement settings 15 Set the acceptance range 15 Set up a LAN connection 16 C...

Page 4: ...gs 27 Do a calibration verification 28 Maintenance 28 Clean spills 29 Clean the instrument 29 Clean a sample vial 29 Clean the vial compartment 30 Troubleshooting 30 Remove air bubbles from the sample...

Page 5: ...EBC Accuracy 2 of reading plus 0 01 NTU from 0 to 40 FNU 10 of reading from 40 to 1000 FNU based on Formazin primary standard at 25 C 77 F Linearity Better than 1 for 0 to 40 NTU on Formazin at 25 C...

Page 6: ...ims such damages to the full extent permitted under applicable law The user is solely responsible to identify critical application risks and install appropriate mechanisms to protect processes during...

Page 7: ...electrocution exists This symbol indicates the need for protective eye wear This symbol indicates a laser device is used in the equipment This symbol identifies a risk of chemical harm and indicates...

Page 8: ...se of doubt contact the manufacturer Safety information for RFID modules W A R N I N G Multiple hazards Do not disassemble the instrument for maintenance If the internal components must be cleaned or...

Page 9: ...ncy energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to...

Page 10: ...rt for LAN connection 4 Power button 9 RFID module indicator optional 5 Power supply connection 10 USB port type A Product components Make sure that all components have been received Refer to Figure 2...

Page 11: ...lation to break down which can result in an electric shock hazard The manufacturer recommends that users with concerns contact technical support Installation guidelines Install the instrument On a lev...

Page 12: ...o not use writing tips of pens or pencils or other sharp objects to make selections on the screen or damage to the screen will occur Refer to Figure 3 for an overview of the home screen Figure 3 Displ...

Page 13: ...ading log calibration log verification log and compare log Refer to Show the recorded data on page 24 Setup Configures the instrument settings Refer to Configure the instrument settings on page 12 Dia...

Page 14: ...mmm yyyy default yyyy mm dd dd mm yyyy or mm dd yyyy Time Format Sets the time format Options 12 or 24 hours default Security Enables or disables password protection for the settings and tasks in the...

Page 15: ...nguage and then push OK Add operator IDs Add a unique operator ID for each person who will measure samples 30 maximum Select an icon operator password and security level for each operator ID 1 Push Lo...

Page 16: ...e barcode with a barcode handset scanner that is connected to the instrument The barcode is added to the sample ID 5 Push OK 6 Select an option Option Description Add Date Time Adds the data and time...

Page 17: ...s the measurement units that show on the display and that are recorded to the data log Options NTU FNU TE F FTU EBC mNTU or mFNU Default FNU Data Log Setup Sets the data log settings Auto Store Measur...

Page 18: ...Select Network then push Setup 5 Select On 6 Push TU5200 IP 7 To get an IP address for the instrument from a DHCP server do the steps that follow a Select Automatic b Push OK two times The IP address...

Page 19: ...network printers that are SNMP enabled and in the same sub network are found by the instrument 1 Push two times then push Setup 2 Push Network Peripherals 3 Select Printer then push Setup 4 Push Netwo...

Page 20: ...th the sample When collecting a sample from a water tap in a distribution system or treatment plant turn the water on for at least five minutes then collect the sample Do not adjust the flow because t...

Page 21: ...l and national regulations N O T I C E Always put a cap on the sample vial to prevent spills in the vial compartment Refer to the illustrated steps that follow to prepare a sample vial for measurement...

Page 22: ...operator RFID tag in front of the RFID module or Push Login Select the applicable operator ID then push Select 2 Select the sample ID as follows Put the sample RFID sticker on the sample bottle in fro...

Page 23: ...cess and lab instrument are calibrated with the same number of calibration points and with the same standards Make sure that the calibrations are not expired Collect a grab sample Collect a 100 mL sam...

Page 24: ...eading from the process instrument shows on the display 11 Prepare a grab sample vial Refer to Prepare a sample vial on page 19 12 Measure the turbidity of the grab sample with the laboratory instrume...

Page 25: ...e laboratory instrument 7 Move the grab sample bottle to the laboratory instrument 8 On the TU5200 push Options Reading Setup 9 Push Bubble Reject then set bubble reject to on 10 If the grab sample is...

Page 26: ...Calibration log Shows the calibration history Verification log Shows the verification history Compare log Shows the recorded comparisons of process and laboratory measurements 1 Push Data Log and sel...

Page 27: ...data point Data points Selects the symbol used for the data points Control Limit Sets the minimum value and maximum value of the readings that show on the graph Show the instrument information 1 Push...

Page 28: ...er with kaolin SDVB 0 1000 FNU 3 point calibration 20 FNU 600 FNU and dilution water with spherical slyrene divinylbenzene Custom Calibration 1 to 6 point calibration 0 02 to 1000 FNU with StablCal or...

Page 29: ...measured 12 Push Store to complete the calibration and record the results to the calibration log 13 To show the calibration log push Options Calibration Log Refer to Show the recorded data on page 24...

Page 30: ...ect 2 Push Verification 3 Clean the verification standard vial with a no lint cloth to remove contamination Keep contamination off the glass Refer to Prevent vial contamination on page 18 4 Dry the ex...

Page 31: ...current safety data sheets MSDS SDS for safety protocols Clean the sample vial when there is contamination in the sample vial after the sample vial is rinsed Items to collect Hydrochloric acid concen...

Page 32: ...micro fiber cloth lint free cloth Liquid oil Cloth water and cleaning agent Troubleshooting Message Solution Startup The self check stopped Clean the vial compartment The vial compartment is dirty Set...

Page 33: ...stable Use the correct calibration standards Invert the standard until no bubbles or large particles show The standard value is out of the measurement range Use the correct calibration standards Inve...

Page 34: ...that the file system FAT32 is installed on the USB flash drive Set the power to off wait 20 seconds and then set the power to on again Look for remaining space on the USB flash drive Set the power to...

Page 35: ...k administrator If the DHCP server is not enabled manually enter an IP address for the instrument Setting IP address failed Contact the network administrator If necessary contact technical support Set...

Page 36: ...ollects on the sample vial during measurement let the sample become slightly warmer before measurement Keep the sample at room temperature or partially put the sample vial into a warm water bath for a...

Page 37: ...60 Hz 1 LZV798 Sample vials with caps 2 LZV946 Vial stand 1 LZV952 Accessories Description Quantity Item no Micro fiber cloth vial cleaning 1 LZY945 RFID tags operator 2 pkg LZQ066 RFID stickers samp...

Page 38: ...36 English...

Page 39: ......

Page 40: ...orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 route de...

Reviews: