4.
Push
Add
.
5.
To enter the IP address of the sc controller, do the steps that follow.
a.
Select IP-Address.
b.
Push
IP-Address
, then enter the IP address of the sc controller.
6.
To enter the host name of the sc controller, do the steps that follow.
a.
Select Hostname.
b.
Push
Hostname
, then enter the server name of the sc controller.
7.
Push
Password
, then enter the password for the sc controller.
8.
Push
OK
>
CLOSE
>
CLOSE
.
9.
Select the channels (parameters) that show on the instrument as follows:
a.
Push
Link2SC
.
b.
Push
Process Monitoring
>
Add
.
c.
Select the channel, then push
OK
.
The current reading and the previous readings for the channel show in graph format.
10.
Do step
9
again to show more channels on the display as necessary.
11.
To change the time interval that shows on the graph, push the graph and select Day, Week or
Month.
12.
To show the current reading for one of the channels in the sidebar menu, select the channel and
then push the minimize icon at the top of the screen.
The current reading for the channel shows in the sidebar menu after the Verification icon.
Measurement
Sample collection
• Collect samples in clean glass or plastic bottles with tight-fitting caps.
• Rinse the container a minimum of three times with the sample.
• When collecting a sample from a water tap in a distribution system or treatment plant, turn the
water on for at least five minutes, then collect the sample. Do not adjust the flow because this can
add particles.
• When collecting a sample from a body of water (e.g., a stream or storage tank), collect at least one
liter (1 quart) and fully mix before taking an aliquot for measurement. If the quality of the sample
source is not constant, collect samples at many locations at different depths as necessary. Then,
mix the samples together to prepare one sample for measurement.
• Fill the container. Let the container overflow with the sample and then immediately put the cap on
the sample container so that there is no headspace (air) above the sample.
• Write the sample information on the container.
• Start analysis as soon as possible to prevent temperature changes, bacteria growth and settling.
Prevent vial contamination
N O T I C E
Do not to touch or scratch the glass of the sample vial. Contamination or scratches on the glass can cause
measurement errors.
The glass must stay clean and have no scratches. Use a no-lint cloth to remove dirt, fingerprints or
particles from the glass. Replace the sample vial when the glass has scratches.
Refer to
Figure 4
to identify where not to touch the sample vial. Always keep the sample vials in the
vial stand to prevent contamination on the bottom of the vial.
18
English