Put the vial in the instrument
C A U T I O N
Personal injury hazard. Never remove covers from the instrument. This is a laser-based instrument and
the user risks injury if exposed to the laser.
C A U T I O N
Personal injury hazard. Do not look into the vial compartment when the instrument is connected to
power.
N O T I C E
Keep the lid closed to keep contamination out of the vial compartment.
1.
Log in to the instrument as follows:
• Put an operator RFID tag in front of the RFID module or
• Push
Login
. Select the applicable operator ID, then push
Select
.
2.
Select the sample ID as follows:
• Put the sample RFID sticker on the sample bottle in front of the RFID module or
• Push
Sample ID
. Select the applicable sample ID, then push
Select
.
Note: To add sample IDs to the instrument, refer to
Add sample IDs
on page 14.
3.
Clean the sample vial with a no-lint cloth to remove contamination.
4.
Dry the external surfaces of the vial with a no-lint cloth. Make sure to dry the bottom of the vial.
5.
Put the sample vial in the vial compartment. Refer to the illustrated steps that follow.
20
English