Troubleshooting
80
10.3 Error codes
describes the error codes.
Table 23 Sensocheck limits for Type A sensors
Slope
Zero
Response time
Cal timer
Adm.
range
25–130 nA
–2–2 nA
720 seconds
maximum
—
35–90 nA
–0.5–0.5 nA
300 seconds
80% expired
30–35 nA
–1.0–0.5 nA
300–600
seconds
80–100%
expired
90–110 nA
0.5–1.0 nA
300–600
seconds
80–100%
expired
< 30 nA
< –1.0 nA
> 600 seconds
timer expired
> 110 nA
> 1.0 nA
> 600 second
timer expired
Table 24 Error codes
Code
Description
Corrective action
ERR 01
Measured value (SAT range)
flashes; value is above or
below the range limit.
Make sure that the correct
sensor is connected to the
transmitter.
Replace sensor and cable.
ERR 02
Measured value (Conc
range) flashes; value is
above or below the range
limit.
Make sure that the correct
sensor is connected to the
transmitter.
Replace sensor and cable.
Summary of Contents for si792 D series
Page 2: ...2 ...
Page 6: ...Table of Contents 4 ...
Page 12: ...Specifications 10 ...
Page 36: ...Interface and navigation 34 ...
Page 44: ...Operation si792 x D 4 20 mA HART 42 ...
Page 66: ...Operation Foundation Fieldbus 64 ...
Page 72: ...Operation Profibus PA 70 ...
Page 80: ...Maintenance 78 ...
Page 86: ...Troubleshooting 84 ...
Page 88: ...Parts and accessories 86 ...
Page 93: ...Index 91 Z zero adjustment 72 ...
Page 94: ...Index 92 ...
Page 95: ...93 ...