Section 10 Troubleshooting
10.1 Sensoface
The Sensoface feature is active whenever Sensocheck is active. This
feature monitors the sensor for defects in the sensor or cable, and
indicates the maintenance status of the sensor (
).
10.2 Sensocheck
Important Note: Sensocheck must be switched off for Type B sensors.
The Sensocheck feature monitors the sensor electrodes and wires for
open circuits. When problems are detected, the error message “Err 33”
is displayed, along with an icon corresponding to the type of problem.
shows the sensor types that Sensocheck monitors.
Sensocheck can be switched off in the configuration menu.
Note: For confirmation a Sensoface is displayed after calibration.
Table 22 Sensoface description
Sensoface
Description
The sensor is functioning properly.
The functionality of the sensor is acceptable.
The functionality of the sensor is not longer acceptable.
Replace the sensor.