Operation—Foundation Fieldbus
49
6.2.2
Standard analog input block (AI)
Three Analog Input Function Blocks provide for cyclic transmission of
measured values (currently measured value with status, alarm limits,
freely selectable process parameter).
6.2.2.1 Operating modes
Use the MODE_BLK parameter to set the following operating modes:
• OOS—out of service. If not write-protected, access to all
parameters is allowed.
• MAN—manual
• Auto—online, normal state
6.2.2.2 Set the parameter and units
Use CHANNEL to set the measured parameter and units (
).
The corresponding measurement unit is selected in the UNITS
subparameter of XD_SCALE (
).
Table 15 Measurement parameters and units
Channel
Parameter
Unit
Unit value
1
Saturation
% (percent)
1342
2
Concentration
mg/l
2001
g/l
2002
ppm
1423
ppb
1424
3
Temperature
°C
1001
°F
1002
4
Zero
A
2003
5
Slope
A
1587
6
Volume
concentration
% percent
1342
ppm
1423
Summary of Contents for si792 D series
Page 2: ...2 ...
Page 6: ...Table of Contents 4 ...
Page 12: ...Specifications 10 ...
Page 36: ...Interface and navigation 34 ...
Page 44: ...Operation si792 x D 4 20 mA HART 42 ...
Page 66: ...Operation Foundation Fieldbus 64 ...
Page 72: ...Operation Profibus PA 70 ...
Page 80: ...Maintenance 78 ...
Page 86: ...Troubleshooting 84 ...
Page 88: ...Parts and accessories 86 ...
Page 93: ...Index 91 Z zero adjustment 72 ...
Page 94: ...Index 92 ...
Page 95: ...93 ...