Operation—Foundation Fieldbus
60
PROCESS_PRESSURE
Sets the process pressure
R/W 4
float
0–9.999 BAR
0–999.9 kPa
0–145 PSI
Default: 1.013 BAR
SALINITY
Sets the salinity value
R/W 4
float
0–45 g/kg
Default: 0 g/kg
Calibration
CALIBRATION_MODE
Sets the calibration mode
R/W 1
uns8
0 = Sat
1 = Conc
CAL_HIGH
The slope of the Oxy sensor in
Ampere
R/W 4
float
30–110 Sensor A
220–525 Sensor B
Default: 60 E-9A
CAL_ZERO
The zero offset resulting from a
calibration in Ampere
R/W 4
float
–2–2
Default: 0 E-9A
CAL_SETTLINGTIME
Shows the settling time in
seconds
R
4
float
Default: 0 s
CAL_RH
Shows the relative humidity in
%
R
4
float
CAL_PRESSURE
Shows the pressure
R
4
float
CAL_TEMP
Shows the temperature which
was used during calibration
R
4
float
Table 19 Transducer block bus parameters (continued)
Parameter
Description
R/W Bytes
Data
type
Range
Summary of Contents for si792 D series
Page 2: ...2 ...
Page 6: ...Table of Contents 4 ...
Page 12: ...Specifications 10 ...
Page 36: ...Interface and navigation 34 ...
Page 44: ...Operation si792 x D 4 20 mA HART 42 ...
Page 66: ...Operation Foundation Fieldbus 64 ...
Page 72: ...Operation Profibus PA 70 ...
Page 80: ...Maintenance 78 ...
Page 86: ...Troubleshooting 84 ...
Page 88: ...Parts and accessories 86 ...
Page 93: ...Index 91 Z zero adjustment 72 ...
Page 94: ...Index 92 ...
Page 95: ...93 ...