56
Maintenance
4.
Leave a 30-cm length of the replacement tubing that extends beyond the end of the
mandrel core and secure the tubing with high-temperature tape.
Note:
The length of tubing is specified in the QuikChem method.
5.
Wrap the tubing around the length of the mandrel core. Leave a 30-cm length of
tubing that extends from the other end and secure with tape.
6.
Install the mandrel on the heater element.
7.
Reconnect the thermocouple plug into the jack on the side of the drip tray.
8.
Connect the tubing to the manifold according to the QuikChem method to be tested.
5.6 Dilutor maintenance
Replace all tubes once a year, as well as the seals of the barrel cavity. Contact the
Service Department for assistance.
When using a diluent other than deionized water (e.g., KCl solution), prime the dilutor
with deionized water at least two times to make sure that the diluent is flushed out
completely.
To maintain a dilutor:
1.
In the Omnion software, go to Configuration>Autosamplers. Press the
PRIME
DILUTOR
button.
2.
If the PDS200 dilutor leaks from the connectors, replace the Teflon tubing. If the
barrel leaks, contact the manufacturer for assistance.
Note:
Always degas the diluent with helium.
3.
Keep all surfaces clean. Use a damp cloth to clean the module surfaces. Dry all
surfaces thoroughly. If an acid spill occurs, neutralize the acid with baking soda and
then proceed to clean all the surfaces.
5.7
Sampler maintenance
The sampler has been designed for low maintenance. No user adjustments are
necessary under normal operating conditions.
5.7.1 ASX-500 Series and ASX-400 Series maintenance
•
Immediately—Wipe off any spills. Neutralize acid spills with a commercial acid
neutralizer solution.
•
Daily—Wipe the sample tray and sampler core unit using a lint-free cloth dampened
with a lab-grade cleaning agent. Do not allow the cleaning agent to come in contact
with the lead screws. Wipe off the cleaning agent using a lint-free towel dampened
with water.
•
Weekly—Wipe loose particles off the y-axis lead screw with a dry, lint-free cloth.
•
Monthly—Replace the pump tubes on the rinse station assembly.
•
Never lubricate the leads screws. The lead screw nuts are compounded with a dry film
lubricant. Oiling the lead screws will cause gumming, galling and binding of the
sample probe assembly.
Summary of Contents for QC8500 Series 2
Page 2: ......
Page 6: ...4 Table of Contents ...
Page 10: ...8 Specifications ...
Page 30: ...28 Installation ...
Page 52: ...50 Operation ...
Page 84: ...82 Troubleshooting ...
Page 88: ...86 Replacement Parts and Accessories ...
Page 90: ...88 ...
Page 92: ...90 Warranty ...
Page 94: ...92 Certification ...
Page 102: ...100 Ion Chromatography ...