Installation
3.3.6 Fluidic connections
3.3.6.1 Connect the rinse station
Rinse solution is pumped into the bottom of the rinse station and drains to a waste
container through drain tubing attached at the top of the station (
).
To connect the rinse station:
1.
Connect the rinse-in end of the wash line assembly to the inlet fitting (bottom).
2.
Install the pump tube with the purple pump tube tabs on the pump. Refer to
.
Note:
The outlet of the tube on the right side of the pump attaches to the inlet of the wash bath
, item 1).
Important Note:
Do not lock down the right side of the pump cartridge until ready to
perform the test.
3.
Drop the other end of the wash line into the rinse solution.
4.
Connect one end of the Tygon drain tubing to the drain fitting (top). Drop the other
end into a waste container.
3.3.6.2 Sample line installation (no dilutor)
To install the sample line:
1.
Connect the end of the sample line with the union fitting to the probe.
2.
Install the pump tubing (green collar locks) in the pump. Refer to
.
Note:
Do not lock down the right end of the pump tubing.
3.
Connect the sample line to the injection valve on the system unit.
Figure 21 Rinse station diagram
1
Drain fitting
5
Inlet fitting
2
Wash bath
6
Rinse-in tubing
3
Mounting block
7
From pump
4
Flow
8
Drain tubing for waste
Summary of Contents for QC8500 Series 2
Page 2: ......
Page 6: ...4 Table of Contents ...
Page 10: ...8 Specifications ...
Page 30: ...28 Installation ...
Page 52: ...50 Operation ...
Page 84: ...82 Troubleshooting ...
Page 88: ...86 Replacement Parts and Accessories ...
Page 90: ...88 ...
Page 92: ...90 Warranty ...
Page 94: ...92 Certification ...
Page 102: ...100 Ion Chromatography ...