Advanced Operations
λ
/ Absorbance formula:
1.
Press
Absorbance Formula
.
2.
The formula selected in the top key determines the number of
wavelength and coefficent keys that will appear below. To
change the absorbance formula, press the top key, select a
formula from the displayed list and press
OK
. When a new
formula is selected, the number of variables below changes to
match.
The following formulas are available:
K
1
A
1
+ K
2
A
2
K
1
A
1
+ K
2
A
2
+ K
3
A
3
K
1
A
1
+ K
2
A
2
+ K
3
A
3
+ K
4
A
4
K
1
A
1
/ K
2
A
2
(K
1
A
1
+ K
2
A
2
) / K
3
A
3
(K
1
A
1
+ K
2
A
2
) / (K
3
A
3
+ K
4
A
4
)
A
1
refers to the absorbance at wavelength 1
A
2
refers to the absorbance at wavelength 2, etc.
K
1
refers to the coefficient at wavelength 1
K
2
refers to the coefficient at wavelength 2, etc.
Coefficients can be set negative where subtraction is required.
Recall Data
Call up saved measurement data or time courses, see
section 5.3 on page 37
.
Instrument Setup
Basic data of the instrument, see
section 5.2 on page 27
.
Table 13 Multi-wavelength setup options (continued)
Options
Description
Summary of Contents for Portable Spectrophotometer DR 2800
Page 2: ...2...
Page 6: ...6 Table of Contents...
Page 8: ...8 Specifications...
Page 22: ...22 Installation...
Page 116: ...116 Maintenance...
Page 118: ...118 Troubleshooting...
Page 120: ...120 Replacement Parts...
Page 122: ...122 Contact Information...
Page 124: ...124 Limited Warranty...