Section 1
Specifications
Specifications are subject to change without notice.
Performance Specifications
Operating Mode
Transmittance (%), Absorbance and Concentration
Source Lamp
Gas-filled Tungsten (visible)
Wavelength Range
340–900 nm
Wavelength Accuracy
± 1.5 nm
Wavelength Reproducibility
< 0.1 nm
Wavelength Resolution
1 nm
Wavelength Calibration
Automatic
Wavelength Selection
Automatic, based on method selection
Spectral Bandwidth
< 8 nm
Photometric measuring range
± 3.0 Ext in Wavelength Range 340–900 nm
Photometric Accuracy
5 mAbs at 0.0–0.5 Abs
1% at 0.50–2.0 Abs
Photometric Linearity
< 0.5%–2 Abs
< = 1% at > 2 Abs
with neutral glass at 546 nm
Stray Light
< 0.2% T @ 340 nm with KV450/3
< 0.1% T @ 340 nm with NaNO
2
Data storage
500 measured values (result, date, time, sample ID, user ID)
User programs
50
Physical and Environmental Specifications
Width
220 mm (8.6 in)
Height
135 mm (5.3 in)
Depth
330 mm (12.9 in)
Weight
4.06 kg (8.95 Ibs) without battery
4.38 kg (9.66 Ibs) with battery
Operating Requirements
10–40 °C (50–104 °F), max. 80% relative humidity (non-condensing)
Storage Requirements
-40–60 °C (-40–140 °F) max. 80% relative humidity (non-condensing)
Additional technical data
Mains connection
15 V- / 30VA
Plug-in power supply unit: (100–240 V/50–60 Hz)
Interfaces
Use only shielded cable with max. length of 3 m.
1 x USB type A
1 x USB type B
Enclosure Rating
IP 41 with closed lid
IP 42 with Protective Cover in place
Safety class
Safety class II
Summary of Contents for Portable Spectrophotometer DR 2800
Page 2: ...2...
Page 6: ...6 Table of Contents...
Page 8: ...8 Specifications...
Page 22: ...22 Installation...
Page 116: ...116 Maintenance...
Page 118: ...118 Troubleshooting...
Page 120: ...120 Replacement Parts...
Page 122: ...122 Contact Information...
Page 124: ...124 Limited Warranty...