Advanced Operations
6.4 Single Wavelength (absorbance, concentration and transmittance
measurements)
The Single Wavelength mode can be used in three ways. For
sample measurements at a single wavelength, the instrument can
be programmed to measure the absorbance, % transmittance or
concentration of the analyte.
Absorbance measures the amount of light absorbed by the sample,
in units of absorbance.
% transmittance measures the percent of the original light that
passes through the sample and reaches the detector.
Turn the concentration factor on to select a specific multiplier for
converting absorbance readings to concentration. In a graph of
concentration versus the absorbance, the concentration factor is
the slope of the line.
6.4.1 Set up single wavelength mode
Press
Single Wavelength
in the Main Menu. Press
Options
for
Parameter Setup.
Table 12 Single wavelength setup options
Options
Description
More
For further Options
Store Off/On
With the
Store On
setting, all measurement data are stored automatically. With the
Store Off
setting, no measurement data are stored.
% Trans/Abs
To
switch to % transmittance, concentration or absorbance readings
λ
Wavelength
To enter the measurement wavelength. Use the alphanumeric keypad to enter the measurement
wavelength. The entered wavelength must be in the range from 340–900 nm.
Timer icon
This functions as a stopwatch. It helps to ensure that the steps of an analysis are correctly timed
(e.g. reaction times, wait times, etc., can be exactly specified). When the specified time has
elapsed, an acoustic signal is emitted. The use of the timer has no influence on the measurement
program.
Concentration Factor
Multiplication factor for converting absorbance values into concentration values.
Concentration
Resolution
To select the position of the decimal point in the calculated concentration readings.
Save as User Program
To store the selected parameters as a User Program, see
section 6.1 on page 61
Recall Data
Call up saved measurement data or time courses, see
section 5.3 on page 37
.
Summary of Contents for Portable Spectrophotometer DR 2800
Page 2: ...2...
Page 6: ...6 Table of Contents...
Page 8: ...8 Specifications...
Page 22: ...22 Installation...
Page 116: ...116 Maintenance...
Page 118: ...118 Troubleshooting...
Page 120: ...120 Replacement Parts...
Page 122: ...122 Contact Information...
Page 124: ...124 Limited Warranty...