Advanced Operations
Peak volume/sample volume methods
1.
Select
Stored Programs
in the main menu. Select the required
program.
2.
Press
Start
.
3.
Analyze a sample without added standard solution in
accordance with the instructions in the Procedures Manual.
When the measurement is complete, leave the sample
cuvette/cell in the cuvette/sample cell holder.
4.
Press
Options>More>Standard Addition
.
An overview of the data of the standard addition procedure is
displayed.
5.
Press
OK
to accept the standard values for standard
concentration, sample volume (total) and standard addition
volume. Press
Edit
to change any of these values.