74
Advanced Operations
Process Timer key
1.
Press the
Process Timer
key to enter any waiting, reaction or
handling times that have to be taken into account. Use the
alphanumeric keypad to enter the time. Press
OK
and confirm
the input by pressing
OK
again.
Note:
This time is integrated into the measurement process.
Note:
The entered measurement sequence is displayed.
[R]
key
/ Reading...
1.
Press the
[R]
key to program a measurement of the substance
that is to be analyzed. Confirm with
OK
.
2.
Press
New
and then
Reading...
and use the alphanumeric
keypad to enter the wavelength at which the measurement is to
be completed. Press
OK
and confirm the input by pressing
OK
again.
3.
If measurements are to be completed at a number of
wavelengths, repeat the above two steps for each wavelength.
Note:
The entered measurement sequence is displayed.
{ }
key
Elements of the measurement sequence that are to be repeated
are placed in brackets.
The left bracket "{"marks the start of the sequence that is to be
repeated and the right bracket "}" marks the end.
Note:
The key showing the right bracket remains inactive until a left bracket
is entered.
1.
Press
{
.
2.
Press the key that defines the sequence that is to be repeated:
[Z]
or
[R]
. Confirm with
OK
.
3.
Press
New
and then press
Zeroing...
or
Reading...
and use
the alphanumeric keypad to enter the wavelength at which the
measurement should be completed. Press
OK
and confirm the
input by pressing
OK
again.
4.
Press
}
to end the sequence.
Note:
If an action such as a zero measurement recurs at different
stages of a measurement sequence, the series of actions is numbered
sequentially (e.g. Z1, Z2, etc.).
Summary of Contents for Portable Spectrophotometer DR 2800
Page 2: ...2...
Page 6: ...6 Table of Contents...
Page 8: ...8 Specifications...
Page 22: ...22 Installation...
Page 116: ...116 Maintenance...
Page 118: ...118 Troubleshooting...
Page 120: ...120 Replacement Parts...
Page 122: ...122 Contact Information...
Page 124: ...124 Limited Warranty...